Product specifications
©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc. Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com
DIFFERENTIAL PRESSURE
95
Model Range One Range Two PDA User Interface
8691 = 869 Pascal P Includes PDA N Standard User Interface
0 to 0.5 ±50 E Expert System
0 to 1 100LD 0 to 100
0 to 5 ±100
0 to 2.5 250LD 0 to 250
0 to 15 ±250
±0.25 500LD 0 to 500
±0.5 ±500
±1 10CLD 0 to 1000
±2.5 ±1000
±5 35CLD 0 to 3500
±15 ±3500
8 6 9 1
Ultra-Low Pressure Generating and Documenting Calibrator
Electrical Connections
transducers, a 2-wire and 4-wire congu-
rable length electropneumatic assembly
is available from 6 to 15 feet.
_
+
EXC
SIG
SIG
COM
_
Current External
Excitation
+
External
Power
Supply
EXC COM SIG
EXC
SIG
SIG
COM
3-Wire External
Excitation
_
+
External
Power
Supply
EXC
SIG
SIG
COM
4-Wire External
Excitation
EXC COM +SIG -SIG
_
+
External
Power
Supply
_
+
EXC
SIG
SIG
COM
Current Internal
Excitation
EXC COM SIG
EXC
SIG
SIG
COM
3-Wire Internal
Excitation
EXC
SIG
SIG
COM
4-Wire Internal
Excitation
EXC COM +SIG -SIG
EXC
SIG
SIG
COM
P-Switch SPST
SPECIFICATIONS
Measurement General Specications Pressure Transducer
Accuracy ±0.04% FS
Pressure Units
(Selectable)
n.W.C., PA, kPa, mbar, cm W.C. Pressure Fittings Barbed, Plug-in O-ring Quick Connects
Precision 0.0002” W.C. Warmup 1 Hour Electrical Banana Plug Jacks
Calibration Stability (Pressure
Span)
0.2% Rdg./yr
Reading Rate 20 Readings/Seconds, Typical Voltage Meter ±0.005% FSO at ±10.5 VDC
Calibration Stability (mA and
Voltage)
0.01% FS/yr
Gravity/Orientation Negligible Current Meter ±0.005% FSO at 4 to 20 mA
Calibration Adjustment Zero Tare
Shock and Vibration 5g, Maximum Excitation 24 VDC Nominal for 4 to 20 mA Output,
Adjustable to 5 to 24 VDC for Voltage Output
Compensated Temperature Range 40°F to 120°F
Communications RS 232
Control
Storage Temperature Range 40°F to 160°F
Display 3.5” Transective Type TFT Color, QVGA, 64-k
Color
Controlled Pressure Stability 0.0002” W.C., Typical
Temperature Eect (Zero) None, Zero, Tare Keypad Pocket PC Touch Pad Minimum Controlled Pressure 0.00005” W.C.
Temperature Eect (Span) 0.01%°F Size: 11” x 14” x 6” (27.9 cm x 35.6 cm x 15.2 cm) Dual Reference Pressure Ranges See Order Info.
Certication NIST Weight 16 lbs. (8.2 kg) Pressure Types Gauge and Dierential
Pressure Media Clean, Dry, Non-Corrosive Gases Overpressure Limit 5 Psid
Power 120/240 AC, 50/60 Hz, Battery Li Ion - 8 Hours
Operation, Integrated Charger
Control Time User Selectable