Product specifications

Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
DIFFERENTIAL PRESSURE
94
Step 1

- Select transducer prole
- Select accuracy specication

Step 2
Real Time

put for testing and calibration
- Apply selected pressure to per-


Step 3
Test Unit
- Return to screen to perform
calibration test sequence
- Review and record results
- Copy and save data into
your calibration database
Simple Pocket PC
User Interface

Easy Step-by-Step User Interface Process

Provides Accuracy and Stability Plots
Handles Multiple Engineering Units


Highest Accuracy to Support Certication of
all Low DP Critical Process Pressure Sensors
True Low Range Dual Reference Pressure Sen-
sors with NIST Traceability
Dual Reference Design Provides Maximum
Accuracy, Repeatability and Resolution
Calibration Capabilities
Analog Pressure Transducers
Pressure Switches
Analog Dial Gauges
Setra Digital Auto-Cal
TM
Products 269 & RPM

Ultra-Low Pressure Generating and Documenting Calibrator

 AC or battery operation - eight hours of operation on full battery
 Rugged, compact carrying case - great for cramped and remote locations
 
harness assembly - for analog transducers, 2, 3 and 4-wire, congurable length
Calibration Management database
 Store and retrieve transducer proles
 Generate as found and as left calibration data
 Print calibration certicates
DESCRIPTION
The Model 869 and 869XP are designed for use in critical environments
that require portability, high accuracy, and periodic low pressure sensor
calibration and documentation to certify regulated processes.
Designed to perform calibration checks on installed pressure transducers,
pressure switches, and gauges, the 869 and 869XP oers users selectable
automated pressure generation proles with up to 101 calibration points.
This NASA patented low pressure generating technology achieves ±0.0002

True dierential pressure is generated with both high and low pressure
ports connected to the unit under test, providing isolation from process
background disturbances. High and low pressure ports are shorted to pro-
duce stable, noise-free zero pressure input, outperforming competitive
active zero pressure systems.
The Model 869XP oers fully automated, hands-o calibration. Its easy-to-
use SMART communications software provides transducer detection and
automated calibration of Setras Model 269 digital transducer and Model

allows the 869XP to simultaneously pressurize the 269 or SRPM under test
and automatically transmit ID and calibration data between the two units.
Model 869
Model 869XP
