Product specifications
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
DIFFERENTIAL PRESSURE
46
Model SRIM
Room Isolation Monitor
FEATURES
  Large 2-Line LCD Display
   Membrane Keypad for Ease of Conguration & 
Menu Navigation
   Upper/Lower Alarm Limit Setting
   Wall & Duct Mount Versions
  Mounting Thickness <50mm
   Easy Calibration without Removing the Wiring or 
Plumbing
  True 2-Wire 4 to 20 mA, 0 to 5 VDC & 0 to 10 VDC, 
Field Selectable
  Unidirectional or Bidirectional Ranges
  Accuracy Options: ±1.0%, ±0.5%, ±0.25% FS
  Ranges as Low as ±0.05 in. W.C.
  CE Compliant 
APPLICATIONS
  Pharmaceutical Manufacturing
 Hospitals
 Laboratories
 Vivariums
DESCRIPTION
In a pharmaceutical, hospital, laboratory, or animal research 
facility, the integrity of the ventilation control system is at the 
be maintained at a positive pressure to prevent product con-
tamination or a negative pressure to prevent contaminants 
from escaping into adjacent areas, the proper pressurization of 
the room is essential.
To assure proper pressurization is maintained in these critical 
environments, a room pressure monitor is employed to measure 
and alert sta and personnel of any change in pressure no mat-
ter how small. A fail-safe solution to monitoring these very low 
pressure changes is Setra’s Model SRIM room pressure monitor, 
which utilizes highly accurate capacitance sensing technology 
to measure and display true low pressure dierential.
SPECIFICATIONS
Performance Data Physical Description Electrical Data
Code F Code H Code C/G Electrical Connection Screw Terminal Circuit 3-Wire (Exc, Out, Com)
Accuracy RSS*
±0.25% FS ±0.5% FS ±1.0% FS
Dimensions See reverse side Output 0 to 5 VDC, 0 to 10 VDC, 4 to 20 mA (Loop Powered)
Non-Linearity (BFSL)
±0.22% FS ±0.49% FS ±0.98% FS
Weight 8.9oz (Duct), 9.8oz (Wall) Excitation 18 to 32 VDC
Hysteresis
±0.1% FS ±0.1% FS ±0.1% FS
Display Custom 2-Line Character LCD Current Consumption 5 mA (voltage output mode)
Non-Repeatability
±0.05% FS ±0.05% FS ±0.05% FS
Pressure Fittings Barbed Fittings for 1/4” Tubing
*Calibrated into a 50K ohm load, operable into a 10K ohm load or greater
Zero Setting Tol.
±0.5% FS ±0.5% FS ±1.0% FS
Case Fire Retardant Plastic UL94V-0
Pressure Media
Span Setting Tol.
±0.5% FS ±0.5% FS ±1.0% FS
Air or non-conductive, non-explosive gases
Thermal Eects
Environmental Data Certications
Compensated Range
±0.06% FS/C
Temperature CE
EN61326-1 & EN61326-2-3 BASIC Immunity & Class B Emission
Overpressure
Up to 10 PSI
 Operating 21.2 to +140°F (-6 to +60ºC) RoHS
*RSS of Non-Linearity, Non-Repeatability & Hysteresis at constant temp.
**Units calibrated at nominal 21ºC. Max thermal error computed from this datum.
 Storage -4 to +185°F (-20 to +85ºC)
Specications Subject to Change without Notice
Operating Humidity 5 to 95% RH (non-condensing)
Duct Mount
ROOM PRESSURE MONITORS