Product specifications
Phone: 800-257-3872 • Fax: 978-264-0292 • setra.com ©2014 Setra Systems, Inc. All rights reserved. The Setra Systems name and logo are registered trademarks of Setra Systems, Inc.
DIFFERENTIAL PRESSURE
24
Multi-Sense® Model 231 
■  
■  
■  -
■  Dual Sensors
■  
■  Hinged Cover
■  
■  Optional LCD Display 
■  All Cast Aluminum, NEMA4 Rated Housing
■  CE and RoHS Compliant
APPLICATIONS
■  Energy Management Systems
■  Process Control Systems
■  -
uids
■  -
sels
■  
NOTE: Setra quality standards are based on ANSI-
The calibration of this product is NIST traceable.
U.S. Patent nos. 6019002; 6014800
DESCRIPTION
 -
sure transducer all-inclusive design provides users with eld 
accessible ranging, choice of output and eld zeroing. 
Choose from three congurable pressure transducers: 5 up 
to 50 psid, 10 up to 100 psid, or 25 up to 250 psid. Each Model 
231 has 4 unidirectional and 4 bidirectional switch selectable 
pressure ranges and can be recongured in the eld for 0-5 
231 jumper selectable port swap feature eliminates costly 
replumbing if the pressure transducer is improperly installed 
or replaced. An optional LCD display is available for on-site 
indication of line and dierential pressure.
SPECIFICATIONS
Electrical Data (Voltage) Performance Data Environmental Data
Circuit 3-Wire
Accuracy RSS
1 
(at constant temp.)  Operating
3
 Temperature °F (°C)  -4 to +185 (-20 to -85)
Excitation
15 to 30 VDC/18 to 30 VAC (Reverse 
Excitation Protected)
Pressure Ranges A, B, C ±1.0% FS Storage Temperature °F (°C)  -4 to +185 (-20 to +85)
Output
4
0 to 5 VDC, 0 to 10 VDC, 1 to 5 VDC
Pressure Ranges D ±2.0% FS Vibration 10g from 50Hz to 2000 Hz
Output Impedance 30 Ohms
Pressure Ranges
Shock 200g
Circuit Consumption
8 mA (typ.) at 5 VDC, 8 mA (typ) at 
10 VDC40 mA (typ.) at 18-30 VAC
A B C D Max. Line Pressure
Physical Description 
MS1 50 25 10 5 50 Case Die Cast Aluminum, Powder Coated
Electrical Data (Current)
MS2 100 50 20 10 100 Pressure Fittings 1/8-18 NPT Internal
Curcuit 2-wire (Reverse Excitation Protected)
MS3 250 125 50 25 250 Electrical Connection 1/2 in. Conduit
Output
5
4 to 20 mA
Pressure Media
Size 4.0 x 6 x 2 in. (102 x 152 x 51 mm)
External Load 0 to 250 Ohms
Liquids or Gases Compatible with 17-4 PH Stainless Steel
Note: Hydrogen not recommended for use with 17-4 PH stainless steel
Weight 1.5 lb
Min. Supply Voltage (VDC)
15 + 0.02 x (Resistance of reciever 
plus line).
1
 RSS of Non-Linearity, Hysteresis, and Non-Repeatability.
2 
from this datum. 
3 
Operating temperature limits of the electronics only. Pressure media 
temperatures may be considerably higher or lower.
4 
Calibrated into a 50K ohm load, operable into a 5000 ohm load or 
greater. 
5 
load. 
Specications subject to change without notice.
Sensor Vacity Volume 0.2 cc
Max. Supply Voltage (VDC)
30 + 0.004(Resistance of reciever 
plus line).
Thermal Eects
2
Compensated Range °F (°C) +32 to +130 (0 to +54)
Zero/Span Shift %FS/100°F (50°C) 2.0 (1.8)
Warm-up Shift <0.12% FS
Response Time 1 to 5 sec. (selectable)
Proof Pressure 2 x Full Scale
Burst Pressure 15 x Full Scale (50 psi), 10 x Full Scale (75 x 
150 psi), 8 x Full Scale (250 psi)