Specifications
Bases Setup • Super II 53
SETUPS
MORE
BASES
SETUP
CALIB
abc reset
tare
menu remote
Base 1
Grams
Select calibration option:
ZERO SPAN BASECOM GO BACK
0.00
Soft Key:
CALIB
Calibration
There are two types of calibration available. The most common type of calibration is referred to as a span
calibration which is a single point readjustment of the calibration curve to a known weight value. The scale
MUST be span calibrated when first installed to compensate for gravity differences between the point of
manufacture and the point of installation. The second type of calibration is called a linearity adjustment.
Multiple weight values are assigned to the calibration curve using several calibration weights.
The scale can be span calibrated as often as desired. A linearity calibration should only be necessary if the
scale is out of tolerance (non-linear) after a span calibration has been performed. Frequency of calibration
is dependent on many factors such as use of the scale, changes in the environment surrounding the scale
and accuracy goals. Consult your local scale dealer for a calibration schedule.
To access the CALIBRATION selection in the Bases Setup menu, follow the menu tree or press the keys in
the order listed below:
Press the CALIB soft key.
Before beginning either calibration procedure, the scale should always be level and in the weighing units
of the calibration weight(s). Without anything on the pan, press the ZERO soft key and the message “Zero
Base Process Activated” should appear on the top line.
SPAN Calibration
1. Press the SPAN soft key to enter the span calibration menu.