Specifications

Appendix D Super II 120
Enter the desired display resolution increment in the selected weight
unit and press ENTER.
Note: The display resolution cannot be greater than 1/50,000 of full
scale capacity.
.
Enter the value of the calibration weight and press ENTER.
Remove all weight from the scale pan and press ENTER.
Wait while a stable reading is acquired.
Place the previously entered calibration weight on the scale pan and
press ENTER.
NO YES ENTER END
Enter calibration weight
and press ENTER
(for best results, calibration weight
should be 25% or more of capacity)
NO YES ENTER END
Enter the display resolution and
press ENTER
NO YES ENTER END
Remove all weight from scale base
and press ENTER
NO YES ENTER END
Acquiring stable weight - please wait
NO YES ENTER END
Place (calibration weight) (units) on scale base
and press ENTER