4.0.  CALIBRATION
  The 265 transducer is factory calibrated and should require no field adjust- 
  ment. Generally, the mounting position will have a zero shift effect on ranges 
  below 1” WC. Whenever possible, any zero and/or span offsets should be 
  corrected by software adjustment in the user’s control system. However, both 
  zero and span adjustments are accessible either on the front of the unit or by 
  removing the optional conduit enclosure. The 265 transducer is calibrated in 
  the vertical position at the factory.
  4.1 Voltage Output Zero Adjustment
  While monitoring the voltage between the positive output (OUT) and  
  common (COM), and with both pressure ports open to atmosphere, the zero 
  may be adjusted by turning the zero adjustment screw. (See Diagram 1 for 
265
Current
Monitoring
Device
9 to 30 VDC
+
_
SPAN 
ADJUSTMENT
ZERO 
ADJUSTMENT
Diagram 1
Diagram 2
- COM
OUT
Power
Supply
Readout
or 
DAS
+ EXC
Model 
265
Voltage Circuit Diagram 
CurrentCircuit Diagram 
  3.2 Current Output Units       
  The Model 265 is a two-wire loop-powered 4 to 20mA current output unit (see   
  Diagram 3). The current flows into + terminal and returns back to the power 
  supply through the - terminal (see Diagram 3). The power supply must be a DC
  voltage source with a  voltage range between 9 and 30 measured between
  terminal + and - terminals. The unit is calibrated at the factory with a 24 VDC
  loop supply voltage and a 250 ohm load.
    +EXC  Connected to positive terminal of DC/AC 
      power supply   
    -COM  Connect as the reference for power 
      supply and output signal
    OUT  Connect to positive terminal of control 
      or pressure 