Manual
Electrical Information for Optional ±15 VDC (Option 622) and ±24VDC (Option 621) 
Excitation Pressure Transducers 
Note: Any gures in parentheses ( ) pertain to the ±24 VDC excitation units only.
Connections
Cable Lead  Function  Note: Other leads in the 
White  +15 (+24) Positive Excitation  cable are not used for 
Black Power Supply Ground  electrical Option 621 and 622 
Violet  - 15 (-24) Negative Excitation  unless the transducer has been 
Yellow  Positive Output  wired for Remote Control (see Remote 
Brown  Negative Output  Instructions)
Shield Case
Electrical - Electrical Option 622 permits operation from a ±15 VDC and electrical Option 621 
permits operation from a ±24 VDC supply (common return grounded), and provides you with 
the negative signal output lead also at the common ground potential. Thus many transducers 
can be operated from one power supply into single-ended loads. Circuit is reversed voltage 
protected for at least 5 minutes. Internal transient suppression network is provided for short 
duration transients to 150 volts.
Grounding - For general usage and best results (lowest noise) grounding of the shield/case 
to the black lead is recommended.
Power Supply Advice - This transducer has been designed to operate from a dual ±15 VDC 
(±24 VDC ) source. Stated another way, the supply is a 30V (48V) center tapped supply. The 
±15 V (±24 V) is 15 V (24 V) above the common power return lead, and the - 15 V (- 24 V) is 15 
V (24 V) below the common power return lead.
All three power leads must be connected. The transducer will not function if just the +15 V 
(+24 V) and 
-15 V (-24 V) leads are connected to the power source.
To avoid circuit damage due to excessive voltage, the positive supply voltage should be held 
within the range of +15 V to +20 V (+22 V to +30 V), and the negative supply voltage should 
be held within the range of -10 V to -20 V (-10 V to -24 V).
Note on long cable use:
In some instances, use of long cables (several hundred feet length) may introduce enough 
cable capacitance into the circuit to cause output oscillations. If encountered, this oscillation 
may be eliminated by connecting a 100 ohm resistor (1/8th watt or larger) in series to each 
of the output leads at the end of the 2 foot transducer cable. These series resistors, of course, 
add to the output resistance.
Electrical Information for 4 to 20 mA Pressure Transducers
Electrical - The circuit is designed to operate as a true two-wire 4 to 20 mA pressure transducer.
Connection
The Setra 4 to 20 mA transducer has two electrical leads. These may be connected for current 
ow in either direction, without change in performance or damage. We suggest that the electri-
cal cable shield be connected to the system loop circuit’s ground, thereby improving electrical 
noise rejection.
4 Adjustments for Zero and Sensitivity
Adjustments for Voltage Output Pressure Transducers (cover must be removed)
Caution: TURN OFF EXCITATION POWER DURING COVER REMOVAL OR REPLACEMENT
Zero Pressure Output - Can be adjusted to zero by potentiometer as shown in diagram. Unit 
factory - adjusted to zero output (±10 mV for Model 204 and ±20 mV for Model 239).
Sensitivity (span) - Can be adjusted by potentiometer as shown on outline diagram. Unit fac-
tory - adjusted to order specications.
Linearity - DO NOT TOUCH - Factory adjusted for best linearity. Touching any adjustments other 
than zero output or sensitivity may necessitate recalibration.
Note: Sensitivity (span) adjustment is not recommended unless a primary pressure standard 
(dead weight tester, etc.) is available for use as a reference.
Adjustments for Current Output Pressure Transducers (cover must be removed)
Zero Pressure Output - Zero output can be adjusted ±0.2 mA about the nominal 4 mA. The zero 
adjustment is the potentiometer shown in the diagram. Changing zero will not change span.
Sensitivity (span) - Sensitivity can be adjusted approximately ±5 mA from the nominal 16 mA 
full scale span. The sensitivity adjustment potentiometer is located as shown in the diagram 
and may be adjusted without aecting linearity calibration.
Linearity - DO NOT TOUCH - Factory adjusted for best linearity. Touching any adjustments other 
than zero output or sensitivity may necessitate recalibration.
Note: Sensitivity (span) adjustment is not recommended unless a primary pressure standard 
(dead weight tester, etc.) is available for use as a reference.
5




