Manual
Installation of Pressure Fittings
Your transducer is designed for most accurate operation when subjected to pressures 
within the designated pressure range. Refer to the catalog bulletin specications for 
proof pressure limits. Subjection to excessive pressure voids the warranty. DO NOT 
OVERPRESSURE.
  Standard sealants such as Teon pipe tape generally are satisfactory. For the most 
sensitive pressure ranges, excessive high torquing of a metal pressure tting may cause 
slight zero shift which may be trimmed out using the zero adjustment. Use of a plastic 
tting often shows no noticeable zero shift. The torquing eect does not appreciably 
aect linearity or sensitivity. The wrench at on the 204 and 239 series should be used 
when installing the positive pressure tting.
A. Positive Pressure Fitting and Media Compatibility
Model 204, C204, 204D:  1/4”-18 NPT internal tting
  For use with gases or liquids compatible with 17-4 PH 
stainless steel. (316 stainless steel optional)
Model 239, C239:  1/8”-27 NPT internal tting
  For use with gases compatible with stainless steel, hard 
anodized 6061 aluminum, Buna N “O” ring. (stainless steel 
in place of aluminum optional)
B. Reference Pressure Fitting and Media Compatibility (Dierential pressure measure-
ments only.)
Model 204D:  1/8”-27 NPT internal tting
  For use with clean dry air, non- corrosive or non-conduc-
tive gases. 
   Maximum line pressure 1000 psig.
Model 239, C239:  1/8”-27 NPT internal tting
  For use with non-corrosive, non-condensing clean dry air 
or other gases. Maximum line pressure 250 psig.
Notes: 
(1)  Dierential pressure Models can be used to measure gauge pressure by leav-
ing the reference port open to atmosphere.
(2)  Line Pressure refers to pressure applied to both high and low sides of the 
sensing device simultaneously. Refer to the product data sheet to determine 
maximum dierential pressure and overpressure limits.
2
High
Low
204D, 239, C239
Installation Instructions for Multiple Hook-Up of 
4-Wire Pressure Transducers (24 VDC Units Only)
PS - Power Supply (nominal 24 VDC)
PT - Pressure Transducer (4-Wire Circuit)
DVM/DAS - Digital Voltmeter or Data Acquisition System
Option #1
Uses a single, ground referenced power supply for excitation and separate, isolated dierential input 
(not ground referenced) readout or data acquisition system for each pressure transducer output.
Option #2
Uses a separate, isolated power supply for each pressure transducer’s excitation and a single, ground 
reference (single ended input) readout or data acquisition system for all of the outputs.
Option #3
Uses a single, ground referenced power supply for excitation and either a single, isolated readout 
with a bipolar switch that “breaks before makes” both the + output and the - output of each pressure 
transducer, or a single data acquisition system with a multiplexer (MUX).
Note: The shield is internally commoned to the case and pressure port of the transducer. When the shield is connected to 
ground, the case and pressure port of the transducer will also be commoned to that ground.
7