Specifications

connector, indicating the location of pin 1.
Installing the Drifter
Personnel
Installation is to be carried out only by suitably qualified personnel.
Installation Environment
The Drifter should be located indoors where it will not be subject to water or other
liquid spills or splashes.
In addition, for the temperature display to be usefully accurate, the Drifter should
be located where it will not be subject to hot or cold drafts, e.g. above a kettle is
an undesirable location from both a moisture and temperature perspective.
Mounting
The Drifter is supplied with a mounting bracket for attaching to the wall.
Mounting screws should not be proud of the bracket by more than 1.8mm
in order to allow the Drifter to fit on the bracket.
The access hole to the wall cavity must be similarly sized to that in the
bracket to provide clearance for the connectors at the back of the Drifter.
See Figure 1 for details.
When wiring is completed and the connectors are mated to the unit, fit the Drifter
as detailed in Figure 2.
If it is ever necessary to remove the Drifter from its mounting bracket, follow the
instructions in Figure 3.
15