ALARM CONDITIONS AND ALARM CODES
Use the table below to decode Eclipse 5 alarm conditions. If other alarm codes are displayed by the Eclipse 5, contact Chart
Technical Support for assistance.
Note: The following table is intended as a guide for the provider, not the user.
CONDITION
ALARM
CODE
ALARM
DISPLAY
GREEN
INDICATOR
YELLOW
INDICATOR
RED
INDICATOR
AUDIBLE
ALARM
WHAT TO DO
“Warming Up. Please
Wait.
004 O2 < 70% ON ON ON OFF
Wait. The system typically takes 3 – 5 minutes to reach
specied performance.
“Warming Up. Please Wait. 008 O2 < 85% ON ON OFF OFF
Wait. The system typically takes 3 – 5 minutes to reach
specied performance.
“3.0” (or ow setting) 000 NO ALARM ON OFF OFF OFF
Nothing. The system is operating properly at the specied
ow rate.
“Low Power Cartridge” 010 LOW BAT ON Flashing OFF 2 Beeps
Plug into external power or replace with a charged Power
Cartridge.
“Warm Power Cartridge” 002 WARM BAT ON Flashing OFF 1 Beep
Plug into external power or replace with a charged Power
Cartridge.
“Low 9V Battery - - OFF OFF OFF
3 Beeps on
POST
Replace the internal 9-volt battery.
“O
2
Concentration < 85%” 008 O2 < 85% OFF Flashing OFF OFF
Perform Annual Filter PM Maintenance. If condition persists,
service is required.
“O
2
Concentration < 70%” 004 O2 < 70% OFF OFF Flashing 3 Beeps
Perform Annual Filter PM Maintenance. If condition persists,
service is required.
“Flow Rate Error / Blocked
Flow
020 FLOWRATE OFF Flashing OFF 1 Beep
Check tubing and/or humidier for obstruction. Check/clean
air intake lter. If condition persists past 10 minutes, service
is required.
“Cannot Charge Power
Cartridge”
001 CHARGER N/A ON OFF OFF
Instruct patient to seek an external power source. Remove
Power Cartridge and allow to cool to room temperature. If
Power Cartridge malfunction persists, service is required.
“One Wire Communica-
tion Loss”
100 ONE WIRE N/A Flashing OFF 1 Beep Check battery connections.
“No Inspiration Detected”
(in Pulse Mode only)
200 P <--> C ON OFF OFF OFF
If no inspiratory eort is detected after 15 seconds, the
Eclipse 5 will switch to Continuous Flow for 15 seconds, and
continue this sequence until a breath is detected.
“Loss of External Power”
(without charged Power
Cartridge installed)
040
********
(ashing)
OFF OFF Flashing ON for 5 min
Instruct patient to seek an alternative oxygen supply
until external power is restored or install a charged Power
Cartridge.
“System Fault” 080 FAIL XX OFF OFF ON
ON 10 seconds
then silent
Reset power, wait 20 seconds, reconnect power, restart
device.
Instruct patient to use back up oxygen supply. Service is
required.