Product Manual

Personal Ambulatory Oxygen System Provider Technical Manual
49
PN 20631679 Rev D
TEST PROCEDURES
PURITY AND FLOW RATE TEST PROCEDURE  PREFERRED METHOD
It is recommended that the Eclipse 5 be tested for oxygen concentration and ow performance. The CAIRE recommended
test setup is shown on the following diagram. Oxygen monitors may or may not have an internal pump to draw samples of
oxygen to be measured. Placement of the oxygen monitor in the test setup depends whether they have an internal pump.
Only one oxygen monitor is needed. An oxygen monitor such as Salter Labs PRO
2
Check Elite
®
or equivalent may be used.
Connect the circuit per the diagram shown below.
Turn the concentrator on and set the target ow on the Eclipse 5 to 3 LPM continuous ow.
Allow the Eclipse 5 to stabilize (can take up to 10 minutes).
Verify the O
2
concentration and O
2
ow rate are within specication (shown in Table 1).
Figure 27: CAIRE Recommended Test Setup to verify performance of Oxygen Concentrator
PURITY AND FLOW RATE TEST PROCEDUREALTERNATE METHOD
Attach a calibrated oxygen monitor to the oxygen outlet port in accordance with oxygen monitor manufacturers
recommendations.
ECLIPSE
Unit
Flow Meter
O
2
Monitor with Pump
(preferred location)
O
2
Monitor
(alternate location)
Vent to
Atmosphere
NOTE:
Some sensing equipment may restrict the actual ow rate and provide inaccurate concentration read-
ings. If testing is performed with the oxygen monitor in the alternate location pictured above, ensure
that the oxygen monitor does not signicantly restrict oxygen ow.