Product Manual

Personal Ambulatory Oxygen System Provider Technical Manual
46
PN 20631679 Rev D
5. Locate the large o-ring on the product tank, replace it with the new large o-Ring included in the PM kit (5022-SEQ). Screw the
new HEPA lter into the carriage holder (wing nut topped housing). Insert and screw the new HEPA lter (wing nut topped housing)
into the carriage holder (product tank) and nger snug in place (do not over-tighten). Re-attach tubing and secure with a cable tie.
Figure 23: O-Ring.
REMOVE AND REPLACE THE COMPRESSOR INTAKE FILTER
Replace the compressor intake lter annually, or more often as needed. The compressor intake lter may become clogged
depending upon the amount of contaminants in the air (smoke, dust, dirt, pollen, etc.) and may need to be changed more
frequently. The most common cause of low concentration and eventual system failure is a dirty or clogged compressor inlet lter.
Environmental conditions usually determine the eective life of a lter.
NOTE:
Always cut the heads of cable ties to avoid damaging the tubing.
CAUTION
Operating the Eclipse 5 with a clogged compressor intake lter may
reduce performance and lead to system damage or premature failure.