Product Manual
Personal Ambulatory Oxygen System Provider Technical Manual
39
PN 20631679 Rev D
ANNUAL MAINTENANCE PROCEDURES
The following section lists procedures that are necessary to maintain the Eclipse 5. Service should only be performed by a
qualied technician. To perform periodic maintenance, the only tools that should be necessary are:
• #1 Phillips Screwdriver
• Wire-cutting pliers
• Small cable ties
• ESD Mat or approved ESD system
Figure 11: Maintenance Tools.
REMOVING THE UNIT COVER
1. Disconnect power supplies and remove Power Cartridge before removing the unit cover.
2. Place the unit horizontally on the front cover.
3. Using a Phillips #1 screwdriver, remove a total of 10 screws (M3x12 Pan Head SEM Screw,
P/N 6974-312-SEQ) from the back of the unit.
• Keep hands out of moving parts
• Disconnect power supplies and remove Power Cartridge before
removing the unit cover.
• ESDSafetyproceduremustbeinplace.
WARNING: