Eclipse 5 Oxygen System
Size H x W x D 19.3 x 12.3 x 7.1 (inches), 49.0 x 31.2 x 18.0 (cm)
Weight Eclipse 15 lbs., Battery 3.4 lbs.
Continuous Flow Settings (measured in LPM) 0.5 to 3.0 LPM (0.5 liter increments)
Pulse Dose Settings (measured in mL) 16-96mL, 128mL, 160mL, 192mL
O
2
concentration 90%+5.5%/-3% @ sea level
O
2
output pressure (nominal) 5.0 psig (34.5 kPa)
Sound level 48 dBA at 3.0 LPM Continuous Flow Mode; 40 dBA at 3.0 Pulse Dose setting
Power AC Power 100-240 VAC, 50-60 Hz
DC Power 12 V nominal
Battery Quantity (2) 97.5 W-Hr battery packs each containing 7.92 gm of lithium content
Power Consumption 52 watts at 1.0 LPM CF, 145 watts at 3.0 LPM, 45 watts at 1.0 pulse dose, 95 watts at 6.0 pulse dose
Battery Duration 2 hours (at 2LPM continuous ow); 5.1 hours (pulse at a 2 setting)
Nominal Battery recharge time 1.8 to 5.0 hrs recharge time to achieve 80% capacity (dependent upon the ow)
Backup Alarm power 9 volt battery
O
2
concentration indicator Green Light=Normal, Yellow=Caution <85%, Red Flashing=Warning <70%
Operating temperature 50ºF to 104ºF (10ºC to 40ºC) at 82.4º F (28ºC) dew point, non-condensing
Trigger Sensitivity Adjustable between -0.135 cm H
2
0 to -0.37 cm H
2
0 (negative pressure)
Altitude Operating range 0-13,123 (0-4,000 meters)
Alarms/Alerts Loss of power, Low Battery, Low Therapeutic O
2
Output, O
2
flow outside normal limits
No Inspiration detected in Pulse Dose Mode, Unit Malfunction
Description Part Number
Eclipse 5 Oxygen System, Model 1000B, North American Power Cord, English Manual 6900-SEQ
NOTE: The Eclipse Oxygen System consists of Concentrator, Power Cartridge (battery),
AC Power Supply, DC Power Supply, Eclipse pack and cart with telescoping handle.
AIRLINE APPROVALS
Eclipse is approved by the Federal Aviation Administration (FAA) for onboard, in-ight use.
CAIRE, Inc.
2200 Airport Industrial Drive, Suite 500
Ball Ground, GA 30107
Ph 770-721-7759 • Toll-Free 1-800-482-2473
Fax 888-932-2473
Chart BioMedical, Ltd.
Unit 6, Ashville Way
Wokingham RG41 2PL, United Kingdom
Ph +44(0) 1189 367060 • Fax +44(0) 1189 799245
Chart Industries reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions,
specifications and/or, processes to its products at any time without prior notice and with no further obligation or consequence.
All rights not expressly stated herein are reserved by us, as applicable.