Product Manual
Personal Ambulatory Oxygen System Provider Technical Manual
85
A Procedure to replace the new Combo Power Manager Board and Battery Bridge Board:
1 Remove and Replace Combo Power Manager board set
1.1 Remove the Power Cartridge and unscrew the 4 screws holding the BBB pcb to the case (PN: 6974-312-SEQ) shown in Fig 63.
Fig 63: Screws
1.2 Remove the Unit cover as described in section, “Remove and Replace the Unit Cover”.
1.3 Disconnect the 5 wire harnesses as shown in Figure 54. Remove the Battery Bridge pcb (BBB pcb) as shown in Fig 63 (Pull straight
up and o of the Power Manager pcb.)
Fig 64: Cable locations
Disconnect power supplies and remove Power Cartridge before
removing the unit cover. Do not touch exposed circuits during
maintenance without ESD protection.
WARNING:
Compressor Cable
Extrenal Power Cable
Cooling Fan Cable
Compressor Cable