Product Manual

Personal Ambulatory Oxygen System Provider Technical Manual
7
Pulse Dose Mode Specications
Pulse Settings 1.0 to 6.0, in 16mL increments and 128mL, 160mL, 192mL
Trigger Sensitivity Adjustable between settings of 1 (most sensitive) to 3 (least sensitive)
Adjustable Bolus Rise Time Adjustable settings of Fast(factory setting), Medium, or Slow
Trigger Criteria
Cannula pressure has dropped below the trigger point (typically between 0.15– 0.45
cm of H
2
O of negative pressure)
At least 1¼ seconds has passed since the last pulse began
Minimum time between
breaths
1.25 seconds (max. 3 consecutive breaths)
Response to Missing Breaths
While in Pulse Dose Mode, the Eclipse 3 is always monitoring for breath detection.
After 15 seconds of no breath detected, the Eclipse 3 delivers” Continuous Flow at the
last Continuous Flow setting. After another 15 seconds, the Eclipse 3 stops delivering
Continuous Flow and waits 15 seconds for a breath. The Eclipse 3 will stay in this
modality until a breath is detected. (See Power Cartridge Conservation Feature, page 26.)
Power Accessory Specications
AC Power Supply
Input Voltage 100-240VAC, 50-60 Hz
Input Power 245-260 VA
Output Voltage 28 VDC
Output Power 200W
DC Power Supply
Input Voltage 11.5-18VDC
Output Voltage 26 VDC
Output Power 150W Max
Power Cartridge (Battery)
Output voltage 14.8 VDC
Capacity
Quantity (2) 97.5 W-hrs batteries
(Each containing 7.92 grams equivalent Lithium content)
Nominal Power Cartridge Life 80% Capacity after 500 Charge/Discharge cycles
Power Cartridge Recharge Time
1.8 to 5.0 hours, dependent on ow setting, to achieve 80% capacity
from a fully discharged Power Cartridge
NOTE: Bolus volume decreases as breath rate exceeds published range.
BOLUS VOLUME +/ 15%
Pulse Dose
Setting
Bolus Size
(± 15%) mL
AC Power Supply and
Power Cartridge (Battery)
DC Power Supply
Max Breath
Rate
Max Breath
Rate
1.0 16 40 40
2.0 32 40 40
3.0 48 40 40
4.0 64 40 31
5.0 80 37 25
6.0 96 31 20
- 128 23 15
- 160 18 12
- 192 15 10