MALFUNCTION CODES
If a malfunction occurs in the Eclipse, the device will stop, the Red LED on the front panel will light and the buzzer will sound for 10
seconds and then silence. The LCD will display one of the following Malfunction Codes:
Step 1: Remove all power sources; battery and external power. Wait 20 seconds. Reconnect power to the Eclipse. Turn the device
to the ON position. If resolved, continue use.
Step 2: If unresolved, Call Chart T/S for assistance.
Malfunction FAIL code Recommended Action
Invalid RESET FAIL 80 Check for secure communication cables, Verify functionality of PM and CS PCBs
I/O Port Failure FAIL 81 Check for secure communication cables, Verify functionality of PM and CS PCBs
RAM Failure FAIL 82 Check for secure communication cables, Verify functionality of PM and CS PCBs
FLASH Failure FAIL 83 Check for secure communication cables, Verify functionality of PM and CS PCBs
EEPROM Failure FAIL 84 Check for secure communication cables, Verify functionality of PM and CS PCBs
IPC Watchdog Timeout FAIL 90 Check for secure communication cables, Verify functionality of PM and CS PCBs
Compressor Motor Too Hot FAIL 91
Allow Eclipse to Cool for 30 Mintues, Check compressor fan functionality, Check to ensure vents
are not obstructed, Verify functionality of the Compressor Assy
PCB Too Hot FAIL 92
Allow Eclipse to cool for 30 Minutes, Verify functionality of PM fan, Check to ensure vents are not
obstructed
Compressor Motor Stalled FAIL 94
Check Compressor wires for damage, Verify functionality of PM PCB, Verify functionality of the
Compressor Assy
Battery Communication Failure FAIL 95
Verify functionality of the Battery Bridge PCB, Check Battery and Battery Bridge PCB for pin dam-
age, Verify functionality of battery, Verify functionality of PM pcb
IPC Failure FAIL A0 Verify functionality of PM PCB
Product Pressure Sensor Failure FAIL A1 Verify functionality of CS PCB
Breath Pressure Sensor Failure FAIL A2 Verify functionality of CS PCB
Ultrasonic Failure FAIL A3 Verify functionality of CS PCB, Check for water intrusion from humidier
Product Temperature Sensor Failure FAIL A4 Verify functionality of DB9 cable harness, Verify functionality of CS PCB
Loss of Product Tank Pressure FAIL A5 Check for leaking, cracked HEPA lter or damaged tubing, Verify ATF motor is turning
Purity Calibration Data Failure FAIL A6 Check for leaking, cracked HEPA lter or damaged tubing, Verify functionality of CS PCB
Flow Calibration Data Failure FAIL A7 Verify functionality of Proportional Valve, Verify functionality of CS PCB
Breath Sensitivity Data Failure FAIL A8 Check for damaged tubing, Verify functionality of CS PCB
Hour of Operation Data Failure FAIL A9 Verify functionality of CS PCB
Invalid Assembly Option FAIL AA Remove Power to Reset
Ambient Pressure Sensor Failure FAIL AB Verify functionality of CS PCB
Unidentied Failure FAIL FF Remove Power to Reset