Product Manual
Personal Ambulatory Oxygen System Provider Technical Manual
48
To replace the compressor intake lter, follow these steps:
1. Cut the cable tie on the silicone tubes attached to the compressor intake lter.
Figure 27: Removing the Compressor Intake Filter.
2. Remove the silicone tubes attached to each end of the lter body.
3. Install a new lter by pushing each tube completely over the barb on the lter body. Verify
ow direction is correct.
4. Reconnect silicone tubes to the lter. Ensure that the feed tubes are not blocked, crimped
or kinked upon completion of the installation or the unit will alarm for low oxygen purity
after the warm-up cycle is complete.
Figure 28: Installing the Compressor Intake Filter.
5. Re-install two cable ties on silicone tubes.