Product Manual

Personal Ambulatory Oxygen System Provider Technical Manual
45
4. Reconnect silicone tubes to the lter. Verify ow direction is correct. Ensure that the feed tubes are not blocked,
crimped or kinked upon completion of the installation or the unit will alarm for no/low oxygen ow after the
warm-up cycle is complete. Route tubing as shown in Figure 22 below. Place HEPA lter into slot in case.
Figure 21: Installing the HEPA Filter.
5. Re-install cable ties on silicone tubes.
NOTE:
Route tubing as shown in Figure 22. Install cable ties loosely as shown.
CAUTION
Once the silicone tubes are disconnected from the HEPA lter the
oxygen circuit is exposed to the atmosphere. Replace and reconnect
the lter immediately in order to minimize exposure to debris and
humidity which may result in ATF Module damage.
Internal Filters (Newer E3s)
HEPA Filter replacement SN: 11F06100456 and above:
1. Cut cable tie and Disconnect the tubing at the top of the lter carriage.
Figure 22: Tubing on Filter Carriage.