eQuinoxTM Model 4000 Oxygen Concentrator eQuinoxTM Modèle 4000 Concentrateur d’oxygène eQuinoxTM Modell 4000 Sauerstoffkonzentrator eQuinoxTM Modelo 4000 Concentrador de oxígeno eQuinoxTM Modello 4000 Concentratore di ossigeno eQuinoxTM Modelo 4000 Concentrador de oxigénio eQuinoxTM Model 4000 Zuurstofconcentrator User Manual Manuel d’utilisation Bedienungsanleitung Manual del usuario Manuale dell’utente Manual do Utilizador Gebruikershandleiding
User Controls & System Status Indicators Symbol A 02 Definition Definition Read user manual before operation. See user manual for instructions. FAA Approved Symbol: The U.S. Federal Aviation Administration (FAA) has approved this device for use on-board commercial aircraft. No Smoking Icon: Do not smoke near unit.
Contents User Controls & System Status Indicators...............................................................................................2 Quick Start Guide.......................................................................................................................................4 Indications for Use.....................................................................................................................................7 Contraindications for Use.................................
Quick Start Guide 1 2 Unpack Your eQuinox Getting to Know Your eQuinox Review all Warnings, Cautions and additional device information in the rest of this manual. Become familiar with the key features of the eQuinox and the User Control Panel.
The eQuinox User Control Panel displays important operating information. ON/OFF Button (Green) Indicator: This button powers the device ON or OFF. The Green Indicator is lit up when the device is ON and functioning properly. Increase or Decrease Flow Setting Buttons: Use these buttons to set the flow to your prescribed setting. 3 Powering on the eQuinox To connect to AC power source: Insert the AC Power Supply cord securely into the External Power Receptacle on the right side of the device.
4 Select Flow Delivery Mode Flow Mode Button and Indicator Press the Flow Mode button to select your desired flow mode. Pressing this button repeatedly will switch you back and forth between continuous flow and pulse dose modes. In Continuous Flow Mode, the oxygen is provided at a constant flow rate between 0.5 and 3.0 LPM, indicated by the letter C. In Pulse Dose Mode, oxygen is supplied in a bolus at the beginning of each inhalation.
Important! Safety Instructions are defined as follows: WARNING: IMPORTANT SAFETY INFORMATION FOR HAZARDS THAT MIGHT CAUSE SERIOUS INJURY. CAUTION: Important information for preventing damage to the eQuinox. Note: Information needing special attention. Indications for Use The eQuinox is intended for the administration of supplemental oxygen. The device is not intended for life support, nor does it provide any patient monitoring capabilities.
WARNING: If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately. WARNING: Operating the oxygen concentrator outside of the operational temperature specifications can limit the concentrator’s ability to meet Oxygen Concentration specification. Refer to the specification section of this manual for temperature limits. WARNING: Use no oil, grease, or petroleum-based or other flammable products with the oxygen-carrying accessories or the Oxygen Concentrator.
Recommended Operating Environments For correct use of the device refer to the following chart. The chart provides important information about the recommended operating environments, or operating conditions. Specifications Operating 32º F to 104º F (0º C to 40º C) Temperature Operating Humidity 10% - 95% at an 82.
Operating Instructions Before Operating General Charging Information This user manual is your reference to help you use and maintain the device. If you have any questions or concerns please call your home care provider. The eQuinox runs on external AC power, DC power, or a rechargeable battery. The system includes a “smart battery charger.” This recharges the battery whenever the eQuinox is connected to an outside power source. The system monitors and controls the power source and the battery charger.
WARNING: Do Not touch the battery contacts and the patient at the same time when energizing the unit with the external power supply (AC Power Supply) and with no battery pack installed. CAUTION: Use only accessories recommended by the manufacturer. Use of any other may be hazardous, cause serious damage to your oxygen concentrator and will void the warranty. CAUTION: Do not use extension cords with this unit or connect too many plugs into the same electrical outlet.
Rechargeable Battery Pack (Battery) Note: Operating the Oxygen Concentrator outside of its normal operating temperature range can affect performance and decrease battery run time and/or increase battery charge time. (Refer to the Specifications section in this manual.) You will achieve longer operating time on the Battery if you operate your eQuinox in the Pulse Mode (Physician approved mode). The eQuinox can be run by the rechargeable Battery Pack (Battery), which is supplied with the device.
Typical Battery Recharge Time Enable/Disable the Voice Module It will normally take between 2 and 5 hours to recharge a fully drained Battery back to 80% capacity, depending on the flow setting being used. When the battery is fully charged, the battery gauge will stop cascading. The battery can be charged in the unit when the eQuinox is turned off and plugged in. If the Battery becomes too warm during discharging, recharging will not begin until the Battery sufficiently cools.
First Time eQuinox Operation Step 1: Positioning Your eQuinox for Use Place the eQuinox in a well-ventilated, well lit area. Be sure the air inlet and exhaust vents in the back are not blocked. Position the eQuinox so that all audible and visual indicators or alarms can be easily seen and heard. Air inlet Filter Be sure the air inlet filter is in place before operating your eQuinox. If the air inlet filter is missing, contact your home care provider.
Pulse Dose Mode and autoSat Features Your eQuinox has a feature called autoSAT . It delivers a consistent-sized bolus of oxygen up to 40 breaths per minute (see Max Breath Rate table below). As your breath rate increases, such as when you walk, the autoSAT feature servocontrols the device, delivering the set bolus size. autoSAT Technology is proprietary to CAIRE and is used during Pulse Mode only.
User Troubleshooting Table Do Not Ignore Alarms WARNING: The Manufacturer recommends an alternate source of supplemental oxygen in the event of a power outage, alarm condition, or mechanical failure. Consult your physician or Equipment Provider for the type of reserve system required. Yellow Light Symptom Possible Cause The yellow light is on solid. The The eQuinox has detected a alarm is giving 1 beep every 30 problem with the connecseconds. tion to the battery pack. The yellow light is flashing.
Red Light Symptom Possible Cause Your Action 1) Ensure the air intake filter is not clogged or restricted. Clean and replace the filter if necessary. The red light is flashing. The alarm is giving 3 beeps every 2 minutes. The eQuinox has detected low oxygen levels 2) Ensure the eQuinox is in a well ventilated area. Make sure there are at least 3 inches between the back of the eQuinox and any obstructions (furniture, curtain, etc.
Other Alarm Conditions Symptom While in pulse mode, the pulse mode LED flashes quickly for 15 seconds. Possible Cause The eQuinox is unable to detect the user’s breathing effort. See page 15 for pulse mode information. Your Action 1) Pulse mode may not work correctly if the user has any condition that causes blocked/restricted nasal passages. (Illness, deviated septum, etc.) 2) Ensure cannula is no longer than 7 feet in length. 3) Ensure that there are no kinks or blockage in cannula tubing.
Cleaning, Care & Routine Maintenance Routine Maintenance Air Inlet Filter Only a service technician trained and qualified by CAIRE should work on the internal parts of the eQuinox. Ambient air is drawn into the device through the air inlet located on the top rear of the device. Cleaning the air inlet filter is the most important maintenance activity you will perform as a user. Cleaning the air inlet filter will keep your eQuinox performing properly.
AC Power Supply, 12VDC Cable, Power Cord and Outer Case: Optional Humidifier Turn OFF the eQuinox and disconnect from AC or DC power before you clean or disinfect. DO NOT spray right onto the outer case. Use a damp (not wet) cloth or sponge. Spray the cloth or sponge with a mild cleaning solution to clean the cabinet and power supply. To disinfect the eQuinox, use Lysol® Brand II disinfectant (or equivalent). Follow the directions of the cleaner manufacturer.
EMC Tables Guidance and Manufacturer’s Declaration—Electromagnetic Emissions The eQuinox is intended for use in the electromagnetic environment specified below. The customer or the user of the eQuinox should assure that it is used in such an environment. Emissions Test RF emissions CISPR 11 RF emissions CISPR 11 Compliance Electromagnetic Environment - Guidance Group 1 The eQuinox uses RF energy only for its internal function.
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity The eQuinox is intended for use in the electromagnetic environment specified below. The customer or the user of the eQuinox should assure that it is used in such an environment. Immunity test IEC 60601 test level Electromagnetic envi±6 kV contact ronment – guidance ±8 kV air IEC 61000-4-2 Compliance level Electromagnetic environment – guidance ±6 kV contact Floors should be wood, concrete or ceramic tile.
Guidance and Manufacturer’s Declaration–Electromagnetic Immunity The eQuinox is intended for use in the electromagnetic environment specified below. The customer or the user of the eQuinox should assure that it is used in such an environment.
Recommended separation distances between portable and mobile RF communications equipment and the eQuinox The eQuinox is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
General Airline Travel Information During taxi, take-off, and landing, the eQuinox must be turned off and stowed under the seat or in another approved stowage location so as to not block the aisle way or the entry way into the row if the eQuinox will not be used.
CAIRE Inc.® 2200 Airport Industrial Drive, Ste. 500 Ball Ground, GA 30107 USA 1-800-482-2473 Worldwide: www.CAIREmedical.com Medical Product Services GmbH Borngasse 20 35619 Braunfels, Germany Worldwide: www.CAIREmedical.com CAIRE Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further obligation or consequence.