eQuinox
Transportable Oxygen Concentrator with autoSAT
®
Product Specifications
Ordering Information
eQuinox Oxygen System
Size H x W x D 13.6 x 10.6 x 7.4 in. (34.5 cm x 26.9 cm x 18.8 cm)
Weight 14 lbs. (6.4 kg) with 12-Cell Battery Pack Installed
Continuous Flow Settings 0.5 to 3.0 LPM (0.5 liter increments)
Pulse Dose Settings 16-96mL (16mL increments), 128mL, 160mL, 192mL
O
2
concentration 90%+5.5%/-3% @ sea level
O
2
output pressure (nominal) 5.0 psig (34.5 kPa)
Sound level 46dBA @ 3.0 LPM Continuous Flow; 45dBA @ 2.0 LPM Continuous Flow;
40dBA @ 0.5 LPM Continuous Flow
Power AC Power 85-264VAC, 47-63Hz
DC Power 12 V nominal
Battery Pack Quantity (1) 89 W-Hr battery pack 12-cell battery; Quantity (2) 95 W-Hr battery packs 24-cell battery
Power Consumption 110 watts at 3.0 LPM Continuous
Battery Duration 1.2 hrs at 2 LPM Continuous; 2.7 hrs at 32mL pulse (Setting 2) (12 BPM) with 12-cell Battery Pack
2.75 hrs at 2 LPM Continuous; 5.9 hrs at 32mL pulse (Setting 2) (12 BPM)
with optional 24-cell Battery Pack
Nominal Battery Recharge Time 1.5 to 3.5 hrs recharge time to achieve 80% capacity (dependent upon the flow)
Backup Alarm power 9 volt battery
O
2
concentration indicator Green Light=Normal, Yellow=Caution <85%, Red Flashing=Warning <70%
Operating temperature 32ºF to 104ºF (0ºC to 40ºC) at 82.4º F (28ºC) dew point, non-condensing
Trigger Sensitivity Adjustable between -0.13 cm H
2
0 to -0.4 cm H
2
0 (negative pressure)
Altitude Operating range -382 to 4000 m International Standard Atmosphere (-1254 ft to 13,130 ft) (1060 to 616 hPa)
Alarms/Alerts Loss of power, Low Battery, Low Therapeutic O
2
Output, O
2
flow outside normal limits
No Inspiration detected in Pulse Dose Mode, Unit Malfunction
Description Part Number
eQuinox Oxygen System consists of Concentrator, Battery Pack, 4807-SEQ
AC Power Supply w/ USA-style cord, DC Power Cable, Accessory Bag and Cart with telescoping handle.
AIRLINE APPROVALS
eQuinox is approved by the Federal Aviation Administration (FAA) for onboard, in-flight use.