3.2.2 Electronic Controller:
1. Hang the controller on the TOP edge of the footboard on the
bed frame.
2. Attach the hose/CPR assembly to the LEFT side of the
controller (See Section 4.5).
3. Attach the LAL hose to the single connector on the LEFT side
of the controller. The LAL hose should always be plugged into
the controller, even when not in use. This will prevent the CPR
assembly from falling to the floor when released and
becoming a trip hazard in an emergency situation.
4. Plug into a grounded 115 Vac 60 Hz electrical outlet.
Controller will turn on automatically and “beep” to alert
caregiver that it is active.
5. To begin operation, press the POWER ON/OFF button (See
Section 4.2)
6. Press Max button for rapid inflation of entire mattress.
7. Always place patient in center of mattress.
NOTE - In the event of a power outage or turning the pump
off, it will always come back on in the mode which was last
selected.
8. Once full, you can begin patient setup.
9. During initial inflation, the integrated bolsters will fill first,
followed by the alternating cells.