7.0 PRODUCT SPECIFICATIONS
7.1 Electronic Controller
Electrical Specifications:
Input Voltage AC
115V
Input Frequency
60 Hz
Current
1.75A
Consumption
<60W
Circuit Protection
Double Fused, 250V, 3.15A
Mode of Operation
Continuous
Protection Against Electric Shock
Class II, Type BF applied part
Performance Specifications:
Zones
2 for Alternating Pressure
Max Flow
25 ~ 40lpm
Max Inflate Pressure
50 ± 2mmHg
Max Inflate Timer
20 minutes
Float Mode Pressure
5 - 25 ± 2 mmHg
Integrated and Extended Side Bolsters
60 ± 2 mmHg
Support Surface Inflation Time
5~10 minutes
Alternate Pressure Cycle Range
4 15 minutes
AP Mode Pressure
8 43 ± 2 mmHg
Patient Weight Range
50-500 pounds (22-226 kilograms)
Mechanical Specifications:
Dimensions, L x W x H
5.5 x 9.4 x10.5 inches
Weight
10.5 lbs
Power Cord
13’ (396 cm) Long
16~18 AWG Hospital Grade
Connection
¼” flow quick couplings
Packaging
1 Piece per Box
Air Filter
1.5 x 1.5 in Polyester Filtering Felt
Environmental Specifications:
Operating Conditions:
Ambient Temperature
40
o
F to 104
o
F
10
o
C to 40
o
C
Relative Humidity
30% to 75% Non-Condensing
Atmospheric Pressure
700 hPa to 1060 hPa
Storage and Shipping Conditions:
Ambient Temperature
-40
o
F to 158
o
F
40
o
C to 70
o
C
Relative Humidity
10% ≥ 100%
Atmospheric Pressure
500 hPa to 1060 hPa
Liquid Ingress Protection
IPX0