Operating instructions

Page 20 of 34
an intermediate level disinfectant, such as ProTech
1
. Bleach and
disinfectant should be used according to the manufacturer’s
instructions. To determine the amount of bleach or disinfectant to
use, determine the amount of water in the washer and then
follow the manufacturer’s dilution instructions. Soak the coverlet
in the disinfectant or bleach during the wash cycle. Rinse
thoroughly in clean water and dry before use.
NOTE! 2.5 ounces of bleach per 10 gallons of water is
approximately 100ppm of available chlorine.
CAUTION! Heat will severely damage the material. DO NOT dry
the coverlet using the “heat” cycle. Air dry, or use a “non-heat
dry cycle, e.g., air fluff.
5.2.2 Washing Alternative
If there are no visible signs of body fluids and/or substances
present, the coverlet can be sanitized between patients. To
sanitize the coverlet:
1. Apply chlorine bleach, or an intermediate level disinfectant at
the appropriate dilution (See Section 5.2.1) to the upper
surface of the coverlet. Bleach/disinfectant may be applied
either by spraying or by hand application.
2. Ensure surface is completely covered with the
bleach/disinfectant.
3. Allow bleach/disinfectant to remain in contact with the surface
according to the manufacturer’s instructions.
4. Remove bleach/disinfectant and rinse.
5. Allow to air dry before use.
5.3 Outside Shell
Wipe down with disinfectant, ensuring that all surfaces come in contact
with the disinfectant. Rinse off with a clean damp cloth and allow to air
dry. Inspect straps and fasteners for signs of wear and tear and replace if
needed.
5.4 Air Cell Assembly
CAUTION! DO NOT machine wash or dry the air cells.
1
ProTech
®
is a tuberculocidal disinfectant cleaner and a registered trademark of Central Solutions,
Inc.