5.1.2 The exterior of the controller and CPR assembly should be
periodically wiped down with a cloth dampened with disinfectant.
CAUTION! DO NOT spray disinfectant directly on the
electrical controller, or immerse the controller in any type of
liquid. This could result in a severe electrical hazard.
5.1.3 Before plugging in the controller, check the power cord for
electrical hazards, e.g., cuts, exposed wires, worn insulation, etc.
If hazards are present, take the controller out of operation
immediately and contact Tridien Customer Service at 800-474-
4225 or 954-340-0500.
5.1.4 To ensure optimal performance, calibration and Bio-Med testing
of your Sentry/ST System, should be performed at least annually.
Contact Tridien Customer Service for calibration and Bio-Med
information. Tridien will provide one free Bio-Med test.
5.1.5 Filter Maintenance
5.1.5.1 The filter is removable and can be changed when
necessary. A spare parts list can be found on page 25 or at
Tridien.com.
5.1.5.2 The filter on the bottom of the controller should be
checked periodically and cleaned as needed. A standard vacuum
hose with a brush attachment should be used to clean the filter.
5.1.5.3 To remove or replace the filter, follow the steps below:
1. Unplug the electronic controller.
2. Remove the filter grill cover and remove the filter. DO NOT
unscrew the filter assembly.
3. Clean the filter by washing in a mild detergent and allow to
air dry.
4. Insert the filter back into the filter housing and replace the
grill cover.
5. If the filter cannot be cleaned or becomes damaged, contact
Tridien Customer Service for replacement filters.
IMPORTANT! Good filter maintenance is critical in keeping your
Sentry/ST controllers in optimal operating condition. Failure to keep the
filter clean will result in system downtime and increased repair costs
5.2 Coverlet
5.2.1 Washing and Disinfecting
If there are visible signs of body fluids and/or substances
present, coverlets should be washed between patients. Coverlets
can be machine-washed using chlorine bleach (50-150ppm) or