Data Sheet

54
10 Common Failure Diagnosis of OTDR product
z Common problems
Problem 1: Blank screen or cannot start up.
Possible reason: Electricity is used up.
Solving method: Charge the battery. Use AD-DC adapter to connect the equipment with
external power supply.
Problem 2: Button cant work properly.
Possible reason: The button is damaged.
Solving method: change the button board
Problem 3: data saving arent allowed.
Possible reason: SD card is used up.
Solving method: Export the curve to empty the memory.
z Optical fiber connection problem
Improper optical fiber connection may lead to trace measurement failure, instructions are as
follows:
Problem 4: if the measured trace is as shown in Fig. 10-1:
Fig.10-1
Possible reasons:
①Using improper optical fiber jumper
joint
The connector isnt clean
The connector is ageing
Solving method:
change the optical fiber
jumper
Clean the connector
Change the connector