User's Manual

Alcohol Breathalyzer
User Manual
SAFE INSTRUCTION FOR USE
Components
How to use
User mode
Contents
iSober S is the high-end alcohol breathalyzer in iSober series.
With the compact body design, it offers comport test for self-tester, workers or drivers.
With one button, iSober S can be operated easily, and users can get accurate and faithful
measurement results with Sentech Korea's advanced sensor technology.
Internal and external memories can be used. Up to 100 test results can be confirmed with
the internal memory.
Bluetooth connectivity is available as an option and iSOBER, a dedicated application,
allows you to see more details.
1. Press the power button to turn on iSober S. (*See more detail from the below note)
2. Warming up until you can see “Blow” sign.
(*Warm-up time is calculated in seconds and is in reverse order.)
3. Blowing through the mouthpiece into the unit for 4 seconds until the sounds concludes.
4. The result is displayed for 10 seconds and iSober S turns off automatically.
(*If you want to test again, please turn on iSober S again.)
5. If you do not perform any operation, iSober S will turn off automatically after 1 minute.
*Note: If you keep press and hold the power switch, you can check the firmware of iSober S,
the last calibration date, and usage count numbers after calibration.
DATA through micro SD card
Specifications
*The above image is what you can get from data thorough micro SD card.
Sensor Type
Battery
Weight
Size (mm)
Warming up time
Time for analysis
Accuracy
Working temperature
Storage
Mouthpiece
Auto power OFF
Duration time for display
Calibration interval
Certificate
Additional Feature
Fuel Cell sensor (*Full size)
AAA Size Alkaline Battery 2EA
82.7g (Including batteries)
105(Height) X 51(Width) X 17(Thickness)
0.000 ~ 4.000 g/l
0.000 ~ 4.000 ‰
0.000 ~ 0.400 %BAC
0.000 ~ 2.000 mg/l
000 ~ 400 mg/100mL
5 sec ~ 120 sec
Within 5 seconds (at 0.050%BAC, 24°C)
±5% (at 0.050%BAC)
0°C ~ 40°C
-10C ~ 50°C
MP1000
After 1 min of waiting time
10 sec
1,000 tests or 12 months whichever comes at first.
CE
Reminder for calibration
Internal memory: 100 test results.
External micro SD card
*Option: BLE 4.1/ compatibility with iSOBER application
Wireless Certificate: CE RED, FCC, Bluetooth SIG
User mode will give you to check iSober S condition in detail.
Please press “Down button” to enter User mode.
Back Back
Please put proper pin code.
If code is failed, please try again.
If you forgot the code, please contact an authorized service center
*Default code: 0000
iSober S, Hard Case, Mouthpieces(x5), AAA Batteries(x2), Pouch, and User manual
*MICRO SD card is not included in the package.
Mouthpiece hole
OLED display
Battery cover
Groove for opening
Micro SD card slot
Power button
Up button
Down button
AAA Battery (Alkaline)
PIN Code Code fail
0 0 0 0 0 0 0 0
Warning Display
How to open the iSober S battery cover
Flow:
The breath sample was not strong enough for measurement.
After warming automatically, please blow strong and steady during sampling.
Bat:
The battery power is low. Please replace the batteries.
CAL:
If you test over 1,000 tests, iSober S will display “CAL” sign for 3 seconds then
warning up for test.
This “CAL” sign will be provided the appropriate time for calibration.
*1,000 tests or 12 months interval whichever comes at first.
Err1:
Pressure sensor error – Please contact an authorized service center.
Err2:
Sensor error – Please contact an authorized service center.
Err3:
Out of operation temperature range – Please check the ambient temperature and
operating temperature again, if the same problem persists, please contact
an authorized service center.
Remove the battery cover by pushing up the groove at the bottom
Warnings
1. Wait at least 20 minutes after eating or drinking before testing. Alcohol remaining in the
mouth, or even excess saliva, may interfere with testing. This includes alcohol-based
mouthwash or similar products.
2. Do not blow smoke, saliva or other contaminants into the mouthpiece, as the sensor
may become damaged.
3. Do NOT use the iSober S as a tool to drink and drive.
4. Keep out of extreme temperatures and out of the reach of children.
5. Please Do Not use manganese battery. Use alkaline battery only.
Warranty
The manufacturer warrants the product to be free from defects in workmanship or
material (excluding calibration) under normal use for twelve (12) months from the date
of purchase.
The manufacturer’s obligation under this warranty is limited to replacing, adjusting or
repairing the unit if returned along with proof of purchase. This warranty is void if the
unit has been tampered with or maliciously damaged.
Calibration & Service
For recalibration and technical support, please contact Sentech Korea or Local service center.
-Manufacturer
Company: Sentech Korea Corp.
Location: 21-6, Jimok-ro 75beon-gil, Paju-si, Gyeonggi-do, 10880, Republic of Korea
Tel: +82-31-8071-4400
e-mail: sentech@sentechkorea.com
-Local service center
Back
-Last CAL
24-08-2016
-All count
-Date&Time
Last CAL
*Check the last calibration date
*Check the usage count
(C: usage number after calibration, F: alcohol reaction, S: total usage number)
Back
-Last CAL
24-08-2016
19:44:03
24-08-2016
-All count
-Date&Time
Date&Time
*Time check
-Last CAL
dd-mm-yyyy-All count
-Date&Time
-Set Date
Set date
*Adjust date
*Adjust date
-All count
-Date&Time
hh:mm:ss
-Set Date 19:47:51
-Set Time
Set time
Back
-Last CAL C: 000035
F: 000009
S: 000500
-All count
-Date&Time
All count
Device Information
Product :Breath Alcohol Tester
Software Version : ver 2.04
Last Calibraion : 10-Mar-2000
Test No Date Time Reading Test Result
52 ‘11-Mar-2000 18:22:57 0.000%BAC PASS
-Date&Time
-Set Date
PIN : 0000
-Set Time
-Set code
Set code
*Adjust pin code
Display
Test result
WAIT
004
23-10-2018, 10:43



BLOW
23-10-2018, 10:43
0.450