User's Manual

AMR Transmitter Mounting Instructions for Invensys (Equimeter, Rockwell)
models R175, R200, R250, R275, R310, R315, R415
5
Remove index
1.
Remove and properly dispose the two (2) screws holding the index onto
the meter, with small flat screwdriver.
2.
Remove the index and carefully clean any debris from it. Inspect for
proper index operation by turning the input shaft and verifying that the
gearing turns freely. Also, inspect the index for any missing or broken
hands. Place the index, or a new index if necessary, on a clean surface
near you for reinstallation later.
3.
With a putty knife, scrap away any gasket material that remained on the
meter when the index cover was removed.