User's Manual
13-Survey & Map Mode Noggin
104
The choices available for Station Interval are:
1) Short
2) Normal
3) Long
4) X-Long
5) Normal x10
6) Normal x20
7) Normal x 40
8) Normal x 50
9) Normal x 100
Each choice listed will be followed by an actual value in meters or inches depending on which
units are selected and which Noggin system is being used. Here is a chart showing the station
interval for each system and setting. Note the calculations for Data per Screen assumes that the
Plot Interval is set to Normal for the particular Noggin system (Noggin 100/250 = 2 pixels per
trace and Noggin 500/1000 = 1 pixel per screen). If this assumption is not true, see the formula
after the charts for calculating this value.
N
OGGIN
100 S
YSTEM
N
OGGIN
250 S
YSTEM
Setting Station Interval Data per Screen
Short 5.0 cm or 1.92 in 16 m or 51.2 ft
Normal 10.0 cm or 3.84 in 32 m or 102.4 ft
Long 20.0 cm or 7.68 in 64 m or 204.8 ft
X-Long 50.0 cm or 19.20 in 160 m or 512 ft
Norm x10 100.0 cm or 38.4 in 320 m or 1024 ft
Norm x20 200.0 cm or 76.8 in 640 m or 2048 ft
Norm x40 400.0 cm or 153.6 in 1280 m or 4096 ft
Norm x50 500.0 cm or 192.0 in 1600 m or 5120 ft
Norm x100 1000.0 cm or 384.0 in 3200 m or 10240 ft
Setting Station Interval Data per Screen
Short 2.5 cm or 0.96 in 8 m or 25.6 ft
Normal 5.0 cm or 1.92 in 16 m or 51.2 ft
Long 10.0 cm or 3.84 in 32 m or 102.4 ft
X-Long 25.0 cm or 9.6 in 80 m or 256 ft
Norm x10 50.0 cm or 19.20 in 160 m or 512 ft
Norm x20 100.0 cm or 38.4 in 320 m or 1024 ft
Norm x40 200.0 cm or 76.8 in 640 m or 2048 ft
Norm x50 250.0 cm or 96.0 in 800 m or 2560 ft
Norm x100 500.0 cm or 192.0 in 1600 m or 5120 ft