4.4 Preparing the EM CPD030 for operation
Function check for the display panel
Switch on main switch, on the right hand side of the front panel after connecting the
power cable.
Check Display Panel by pressing the DISPLAY CONTROL key.
All 6 LED symbols must light up while the button is being pressed.
Function check for the magnetic stirrer
Insert the magnetic stirrer (included in the set of accessories) into the specimen
pressure chamber and press the STIRRER key.
LED STIRRER lights up and the inserted magnetic stirrer rotates. Switch off stirrer
by pressing the STIRRER key again.
Setting "Upper" Nominal Temperature
Press the SET TEMP key.
The factory pre-set "upper" nominal temperature of 40°C appears on the
temperature display while the SET TEMP key is being pressed.
If adjustment of the upper nominal temperature is needed, turn the HEATING
potentiometer located at the rear panel of the unit left or right with a small
screwdriver. Turning right increases and turning left decreases the upper nominal
temperature.
"Upper” nominal temperature for CO2 as transition liquid is 40
q
C.