Instruction manual

Page 17 Leica EM CPD030 Operating Manual 09/09
Electrical Connection
Check assigned connecting voltage for the unit of the mains connection and, if
necessary, adjust required connecting voltage, observing, the following procedure:
Push cover (2) by inserting e.g. a screw-driver into the upper slot (3).
Remove voltage selector (4) and reinsert so that the desired connecting voltage is
shown at the display (1).
Gently pull out the fuse holder (5) and insert fuses with the required value: 4 AT for
220 V respectively 5 AT for 115 V.
Reinsert fuse holder (5) according to the arrows indicated and close cover (2).
Connecting the power cable:
Fit locally required mains plug to power cable, considering the following figure of the
power cable socket (L I + N + PE).
Connect power cable to wall socket.
Pull here