Instruction manual
Page 8 Leica EM CPD030 Operating Manual 09/09
Connection data:
Electrical:
Voltage 230 V / 240 V / 115 V
Frequency 50 / 60 Hz
Power consumption 220 VA
Main fuse F1 / F2 4 A slow blowing (230 V / 240 V)
5 A slow blowing (115 V)
Transitional fluid:
Inlet M 12 x 1.5
Outlet Ø 6 mm (R1/8")
Gas outlet Ø 6 mm (R1/8")
Highest permissible pressure
Safety bursting membrane approx. 150 bar
Refrigerator
Cooling capacity 85W
Refrigerant R12
Quantity 120 g
Temperature measurement
COOLING range adjustable 2°C to 12°C
HEATING range adjustable 30°C to 45°C
Operating parameters
Cool down time approx. 2.0°C/min.
Heating time NORMAL approx. 3.2°C/min.
Heating time SLOW approx. 0.9°C/min.
Operating requirements
Carbon Dioxide (chemical formula C02)
Minimum 0,5 kg per maximum run
C02 pressure bottle with feed pipe
Transitional fluid: Acetone