YSI Model 5000 YSI Model 5100 Dissolved Oxygen Temperature Operations Manual
Table of Contents 1. General Description ..............................................................................................................................1 2. Specifications........................................................................................................................................3 3. Quick Start .............................................................................................................................................5 4. Controls .................
11. Maintenance..................................................................................................................................52 Batteries ........................................................................................................................................................ 52 12. Troubleshooting ...........................................................................................................................54 Error Messages ..................................
1. General Description Model 5000 Features The Model 5000 is a microprocessor based, menu-driven, dissolved oxygen meter designed to perform laboratory measurement of dissolved oxygen and Biochemical Oxygen Demand (BOD). The instrument’s menu system makes it simple to use. The Model 5000 has a new case design which facilitates laboratory measurement. The angled profile makes these instruments both functional and attractive.
Model 5100 Features The YSI Model 5100 has all of the same functionality of the Model 5000 and much more. The YSI Model 5100 is a state-of-the-art, microprocessor-based, dissolved oxygen instrument with many automated and application specific features. The YSI 5100 DO instrument contains built-in application software for the calculation of Oxygen Uptake Rate (OUR) and Specific Oxygen Uptake Rate (SOUR).
2. Specifications Oxygen Measurement Ranges: mg/L: 0.0 to 60.0 mg/L % air saturation: 0.0 to 600.0% air saturation mbar: 0 to 1500 Accuracy: mg/L: ±0.1% plus 1 Least Significant Digit (LSD) % air saturation: ±0.1% plus 1 LSD mbar: ±1% plus 1 LSD Resolution: mg/L: 0.1%, or 0.01 mg/L , whichever is greater % air saturation: 0.1% air saturation mbar: 1 mbar Temperature Measurement Range: -5.0 to +50.0°C Accuracy: ±0.1°C Resolution: 0.
Salinity Compensation Range: 0.0 to 40.0 ppt Accuracy: ±.02 mg/L Operating Environment 0 to 45°C, 10 to 90% relative humidity, non-condensing Water Resistance The Model 5000 and 5100 are designed exclusively for indoor use and are NOT waterproof. Power The Model 5000 & 5100 are powered by an AC adapter or 4 C-size alkaline batteries. A new set of alkaline batteries will power the selected instrument for approximately 30 hours (not including stirring). Size and Weight 22.9 by 24.1 by 11.2 cm; 1.
3. Quick Start When you unpack your new Model 5000 or 5100 there are several things you’ll want to do to set the instrument up. 1. Inspect the product to assure that it has not been damaged during shipment. 2. Check the packing list to make sure you have received everything you should have. 3. Install the batteries (see 11. Maintenance, Batteries). 4. Plug the power supply into its mating connector on the back of the meter (see 4.3 Rear Panel). 5.
4. Controls 4.1 Front Panel The front panel of the instrument contains the display and keypad as shown below. The front panel display and controls of the 5000 and 5100 are identical.
The 5000/5100 has two main operating modes, Main and Application. The current mode is displayed in the dark bar at the top right corner of the display. The following diagram shows the display while the instrument is in the Main mode. Mode * 8.15 Main mg /L 11:33AM 24.38°C STORE REVIEW SEND CALIBRATE Soft-key functions The bottom row of the display identifies the function of the four soft-keys which are located on the keypad below the display.
5000/5100 Software Flow Chart MAIN APPLICATION OUR Store Start Review Recall End Send All Send on/off Clear Setup Delete All Review Send Next Rec. Calibrate Modify Auto Cal Send DO Cal Send Setup SOUR Display Start Size End Unit Send on/off Contrast Setup Last Digit Review Report Next Rec. SDF Modify CDF Range Send Next Send Stable Remote System Skip Audio On/Off Up Time/Date Down Version Confirm RS-232 NOTE: Shaded boxes are only available on the model 5100.
4.3 Rear Panel The rear panel contains the connections for the power supply, probe, bar code reader, external computer keyboard and RS232 serial port as shown below. The bar code reader and external computer keyboard share the same socket and are only available on the model 5100. Power Supply Connection The power supply connection requires a 4.5 to 5.5 VDC power supply with at least 350 mA current (6.0 VDC 800 mA when using the YSI 5015 bar code reader).
Bar Code Reader Connection The YSI 5015 bar code reader connects to this 6-pin mini DIN connector. The connector is marked with an arrow to show proper alignment. Be sure to align the arrows when plugging in the YSI 5015 bar code reader (see 7.2 Application Mode, Remote, Bar Code Reader). The bar code connector is also used for attaching an external computer keyboard. Using a keyboard simplifies data entry when using the 5100 in the remote mode, such as with YSI BOD Analyst software (see 7.
5. Setup Before operating the 5000/5100 for the first time, you should set the instrument parameters. Once set, these parameters are maintained in memory. These parameters are discussed below. From Main mode, press the [CALIBRATE] soft-key to enter the calibration menu. Notice that the dark bar in the top right corner of the display shows the current mode. 8.15 mg/L 740 mmHg 11:33AM 98.0 % 0.0 ppt 24.
5.1 Display Setup Press the [DISPLAY] soft-key to show the following menu. * 8.15 Display mg /L 11:33AM 24.38°C SIZE UNITS CONTRAST LAST DIG. The [SIZE] soft-key cycles among the two possible display arrangements for Main mode. One selection shows as many parameters as possible, the other shows Dissolved Oxygen in large type with temperature and time in small type. These different display arrangements are for Main mode only. The display will show all parameters in Calibration mode. 8.
DISPLAY UNIT SETUP D.O. [ % ] Barometer [ in Hg ] Salinity [ Sal – 78 ] Temperature [ C ] Date Format [ MM/DD/YY ] UP DOWN Units HOUR [12H] NEXT The [CONTRAST] soft-key allows the user to change the display contrast. The display contrast will change slightly each time the [UP] or [DOWN] soft-key is pressed. The [LAST DIG.] soft-key allows the user to suppress the last digit of the DO reading. This function is only available when the large size display has been selected.
5.2 Report Setup From the Setup menu, press [REPORT] to display the following soft-keys. [SDF] Space Delimited Format [*] SAMPLE ID # [*] DATE [*] DO% [*] TIME [*] DO mg/L [*] BAROMETER [*] TEMPERATURE [*] SALINITY SEND Record from [ 0] to [99] SDF CDF RANGE Report NEXT The display lists the parameters that can be included in the output string at the RS232 serial port, the default set has everything selected for inclusion.
Comma Delimited Format "SAMPLE ID","mg/L","%","C","ppt","mmHg","TIME","DATE" "ID: 0", 7.95, 97.3,25.6, 0.0, 786,"15:06:34","01/23/96" "ID: 1", 7.94, 97.1,25.6, 0.0, 786,"15:06:36","01/23/96" "ID: 2", 7.95, 97.2,25.6, 0.0, 785,"15:06:44","01/23/96" The asterisk character in front of each parameter indicates that the parameter will be included in the report. Press the [NEXT] soft-key to select the parameter that you want to change. Then press [ENTER] to turn the asterisk character on or off.
5.3 Autostable Setup The autostable feature indicates when readings are stable by emitting a single beep and displaying an asterisk (*) to the right of the dissolved oxygen reading. The instrument uses criteria that you input to determine what a stable reading is. You select the maximum percent of change that may occur during a selected time duration. The default values are 10 seconds and 0.4% variation. The instrument will display an asterisk (*) only when these criteria are met.
5.4 System Setup From the Setup menu, press the [SYSTEM] soft-key to display the following menu. * 8.38 mg /L System 11:47AM 24.38°C AUDIO off TIME / DATE VERSION RS232 The [AUDIO on/off] soft-key allows setting the beeper on or off. The key toggles between an [AUDIO off] soft-key and an [AUDIO on] soft-key. The present state of the beeper would be the OPPOSITE of the function shown on the soft-key. If the beeper is currently on, the soft-key shows “AUDIO off”; i.e., your alternate choice.
* 8.38 mg /L System 9:59AM 24.38°C TIME SETTING AUDIO off SAVED Message TIME / DATE VERSION RS232 If you wish to return to Main mode, press [MODE] three times. The [VERSION] soft-key displays the instrument software version on the message line of the display just above the soft-key descriptions. The [RS232] soft-key displays the RS232 setup. [ 2 ] BAUD RATE 19200 0. 1. 2. 19200 3. 9600 4. 4800 5. 2400 6. 1200 UP RS232 DOWN Select the baud rate by pressing the [UP] or [DOWN] soft-keys.
6. Calibration From the Main mode, press the [CALIBRATE] soft-key to enter Calibration mode. The bottom row of the display shows the soft-key menu. 98.0*% 0.0 ppt 24.38°C AUTO CAL 8.15 mg/L 740 mmHg 11:33AM DO CAL SETUP Calibrate DIAGNOSIS 6.1 Dissolved Oxygen Calibration Dissolved oxygen calibration can be done automatically with the push of one button (see Auto Cal) or manually by entering the desired value in milligrams per liter or percent saturation (see DO Cal).
Auto Cal 1. Prepare the probe according to the instructions in the Probe Operations Manual. 2. Connect the probe to the meter. 3. Place the probe in a BOD bottle containing about 1″ of water to provide a 100% relative humidity environment. 4. Press [ ] to turn the instrument on. 5. Allow the probe to polarize and the temperature to stabilize for at least 15 minutes. If calibration is performed prematurely the values will drift and may be out of specification. 6.
DO Cal (manual) 1. Prepare the probe according to the probe instructions. 2. Connect the probe to the meter. 3. Place the probe in a known oxygen environment, such as a BOD bottle containing about 1″ of water to provide a 100% relative humidity environment or a Winkler-titrated sample. 4. Press [ ] to turn the instrument on. 5. Allow the probe to polarize and the temperature to stabilize for at least 15 minutes. If calibration is performed prematurely the values will drift and may be out of specification.
Flashing digit 98.0*% 0.0 ppt 24.38°C UP 08.15 mg/L 740 mmHg 11:33AM DOWN DIGIT Cal mg/L NEXT 9. Press [ENTER] to confirm your calibration. The screen will momentarily display “D.O. CALIBRATION SAVED” as shown below. 98.0*% 0.0 ppt 24.38°C 8.15 mg/L 740 mmHg 11:33AM D. O. CALIBRATION UP Calibration Message SAVED DOWN DIGIT NEXT NOTE: If you wish to abort calibration before you have pressed [ENTER], you may press [MODE] to return to the calibrate menu without saving the new calibration value.
Flashing digit 98.0*% 0.0 ppt 24.38°C 8.15 mg/L 0740 mmHg 11:33AM Press ENTER to SAVE UP Barometer SETTING DOWN DIGIT NEXT Using the [UP], [DOWN] and [DIGIT] soft-keys, enter the true local barometric pressure. This corresponds to a reading from a mercury barometer. Do NOT use the pressure reported by the weather bureau. Weather bureaus correct pressures to sea level. NOTE: You may estimate the standard pressure at your altitude by using Appendix B - Pressures and Altitudes.
Flashing digit 98.0* % 0.0 ppt 24.55°C 8.15 mg/L 740 mmHg 11:35AM Press ENTER to SAVE UP Salinity SETTING DOWN DIGIT NEXT Using the [UP], [DOWN] and [DIGIT] soft-keys, enter the salinity value. Press [ENTER] to confirm. The message “SALINITY SETTING SAVED” will be displayed as shown below. 98.0* % 10.0 ppt 24.57°C 8.
A standard method for creating such an environment is to dissolve excess sodium sulfite (Na2SO3) and a trace of cobalt chloride (CoCl2) in water. Preferably, the water should come from the sample to be measured. These chemicals will remove all oxygen from the sample (See Standard Methods for the Examination of Water & Wastewater, method 4500-O G, 19th edition). Alternatively, you may place the probe in 100% nitrogen gas. 1.
7. Operation 7.1 Main Mode When the instrument is turned on it is in the Main mode. The following chart shows the display during Main mode operation. * 8.15 Main mg /L 12:13AM 24.38°C STORE REVIEW SEND CALIBRATE The bottom row of the display shows the functions of the four soft-keys used during Main mode. Main mode is used to make measurements, store or review readings and send the current readings to a computer or serial printer.
Store Press the [STORE] soft-key to enter Store mode and display the following menu. * 8.15 Store mg /L Record: 01 11:33AM 24.38°C UP Record number DOWN DIGIT SAVE The Store menu is used to control the storing of data. The current record number is shown in the display on the right side. Press the [SAVE] soft-key to store the values currently displayed.
If the current memory location already contains data and you press [SAVE], the current readings will be stored and the previous readings will be erased. NOTE: There is NO way to restore previous data once it has been deleted. When you have finished storing records, press [MODE] to return to Main mode. Review Press the [REVIEW] soft-key to enter Review mode and display the following menu. 8.15 mg/L 740 mmHg 11:33AM 98.0 % 0.0 ppt 24.38°C RECALL SEND ALL CLEAR Review Record: 01 Record Full 02/06 DEL.
Send All From the Review menu, you may send ALL stored data to a computer or serial printer by pressing the [SEND ALL] soft-key. See 5.2 Report Setup for format. Clear From the Review menu, press the [CLEAR] soft-key to enter Clear mode and display the following menu. 98.0 % 8.15 mg/L Clear 0.0 ppt 740 mmHg Record: 01 Record Full 11:33AM 24.38°C UP DOWN Record number 02/06 DIGIT DELETE The Clear menu is used to delete data that was previously stored.
Delete All To delete ALL stored data, press the [DEL. ALL] soft-key to display the following: 98.0 % 8.15 mg/L Delete All 0.0 ppt 740 mmHg Record: 01 Record Full 11:33AM 24.38°C Press ENTER to DELETE RECALL Record number 02/06 ALL SEND ALL CLEAR DEL. ALL Press [ENTER] to confirm your decision and delete ALL stored data. Press [MODE], instead of [ENTER], if you wish to abort. NOTE: There is NO way to restore data once it has been deleted. Press [MODE] to return to Main mode.
Comma Delimited Format "mg/L","%","C","ppt","mmHg","TIME","DATE" 12.19,138.2,21.6, 0.0, 790,"10:20:56","02/06/96" NOTE: If you want to send stored records, you must first enter Recall mode (see 7.1 Main Mode, Review, Recall).
7.2 Application Mode The YSI Model 5100 DO instrument contains built-in application software for the calculation of Oxygen Uptake Rate (OUR) and Specific Oxygen Uptake Rate (SOUR). These application features may simplify compliance with USEPA 40 CFR Part 503 regulations for vector attraction and provide useful tools for plant operation decisions. With the Model 5100 complete sets of data from five OUR tests and five SOUR tests can be stored to memory and later reviewed.
From the Model 5100 Application mode menu press the [OUR] (Oxygen Uptake Rate) soft-key. The following screen will be displayed. 4.7 23.21°C START OUR mg /L /h 8.92 SETUP mg /L REVIEW SEND: off 0:00 SEND Press [SETUP] to change the OUR parameters. The following screen shows the default parameters. PARAMETERS Sample / Total Min. Time Max. Time Min. Beginning DO Min. Ending DO UP OUR 1/ [ [ [ [ [ 1] 1] min. 15] min. 5.00] mg/L 2.
2.6 OUR mg /L /h 23.21°C 8.92 mg /L SEND: off 0:00 RS232 status Elapsed Time OUR SETTING SAVED START SETUP REVIEW SEND Once the parameters have been set, you are ready to begin. Place the probe in the prepared sample and make certain no air bubbles are trapped, then turn on the stirring and wait a few seconds for the temperature readings to stabilize. NOTE: It is normal for the OUR reading to oscillate for the first few seconds. Press [START] to begin the OUR measurement.
If you want to stop the OUR measurement before the maximum time or minimum ending DO (entered in OUR setup) is reached, press [END]. The final OUR reading will be displayed on the screen. OUR 50.5 Mg /L /h 21.21°C 0.54 mg /L SEND: on 14:20 PROCESS COMPLETE START SETUP STORE SEND NOTE: The DO and temperature readings do not stop updating on the screen.
If you want to save these values press the [STORE] soft-key. A record number will then show up under the REVIEW menu. If the record slot contains a set of data, an indicator saying FULL will appear beside the test number. If the record slot does not contain a set of data, the indicator will say EMPTY. 62.9 mg /L /hr 8.5 mg /L REVIEW Record Number REC:1 EMPTY mg /L 870 0 NEXT REC. MODIFY SAVE 11:04 AM 09/21/98 FULL or EMPTY indicator SEND Press the [NEXT REC.
SOUR The 5100 uses the following formula to calculate the SOUR value: OUR = DOSTART - DOEND TELAPSED x 3600 Sec Total Volume = mg/L/h x 1 Hour Sample Volume Where: DOSTART = Dissolved oxygen level at start of test DOEND = Dissolved oxygen level at end of test TELAPSED = Elapsed time of test in seconds Total Volume Sample Volume SOUR = = Dilution factor of sample (entered in setup screen as Sample/Total) OUR Solids Weight = mg/h/g (mg O2 / hour / g solids) Where: Solids Weight = Total Solids or
Press [SETUP] to change the SOUR parameters. The following screen shows the default parameters. Sample / Total Min. Time Max. Time Min. Beginning DO Min. Ending DO Solids Weight UP 1/ [ 1] [ 1] min. [ 15] min. [ 5.00] mg/L [ 2.00] mg/L [ 1.000] g/L DOWN DIGIT SOUR [*]SOUR@20 SEND: off NEXT Use the [UP], [DOWN], [DIGIT] and [NEXT] soft-keys to change the following parameters as necessary: Sample / Total: Enter the ratio of sample volume to total volume.
7.7 SOUR mg /h /g 25.35°C 8.09 SEND: off mg /L 0:00 RS232 status SOUR SETTING SAVED START SETUP REVIEW SEND Once the parameters have been set, you are ready to begin. Place the probe in the prepared sample and make certain no air bubbles are trapped, then turn on the stirring and wait a few seconds for the temperature readings to stabilize. NOTE: It is normal for the SOUR reading to oscillate for the first few seconds. Press [START] to begin the SOUR measurement.
If you want to stop the SOUR measurement before the maximum time or minimum ending DO (entered in SOUR setup) is reached, press the [END] soft-key. The final SOUR reading will be displayed on the screen. 20.07 @20 mg /h /g 25.33°C 1.29 mg /L SOUR SEND on 11:38 PROCESS COMPLETE START SETUP STORE SEND NOTE: The DO and temperature readings do not stop updating on the screen.
20.1 mg /h /g @20 6.8 mg /L REVIEW Record Number REC:1 EMPTY mg /L 705 0 NEXT REC. MODIFY SAVE 10:32 AM 09/21/98 FULL or EMPTY indicator SEND Press the [NEXT REC.] soft-key to change the record number to find an empty slot, or pick a slot to overwrite (you cannot delete results, you must overwrite). Press the [SAVE] soft-key to save the current test data into that slot. Once you have saved the data you can press the [MODIFY] soft-key to change the parameters of the test. SOUR Time Reference 20.
Remote From the Model 5000/5100 Application mode menu press the [REMOTE] soft-key. The following screen will be displayed. REMOTE STIR ON SKIP 96.4% 24.94 C ID: 1680 MODE :0 UP DOWN CONFIRM The Remote mode is used in conjunction with YSI 5910 BOD Analyst software to greatly reduce the labor required for processing BODs and calculating BOD values. Remote mode is also used to control the 5000/5100 via computer. This can be done using any standard communications program.
1. Prepare the scanner by installing the ferrite bead on the scanner’s cable near to where it plugs into the 5100 meter as shown in the following diagram: 2. Turn to page 1-7 and scan the “Keyboard Wedge Interface for IBM PC AT and compatibles” bar code label. The scanner should beep, indicating that it is ready to be used. 3. Test the bar code reader configuration by selecting Remote mode from the Application menu of the 5100 and scanning a BOD bottle label.
3. Turn to the page entitled “Terminal Selection”, the first page in the booklet. 3.1. Scan the ENTER TERMINAL SELECTION bar code. 3.2. Scan the 0 code on the “Bar codes for menus” page that you unfolded earlier. Hold the reader close to the page to prevent scanning more than one code at a time. 3.3. Scan the 1 code on the “Bar codes for menus” page. 3.4. Go back to the “Terminal Selection” page and scan the EXIT TERMINAL SELECTION bar code. 4.
12. Troubleshooting for additional information on error codes. From the Main mode, press the [CALIBRATE] soft-key to enter Calibration mode, then press the [Diagnosis] soft-key to enter Diagnosis mode. The following screen will be displayed. Diagnosis HISTORY SENSOR HELP RESET Ps History Press the [HISTORY] soft-key to display the following information. Date 02/01 02/02 02/05 02/06 02/07 Time 14:12 08:42 09:55 10:58 13:08 HISTORY DO uA 16.8 16.3 17.0 17.1 16.9 %/uA 5.98 7.90 6.42 7.47 6.
7.3 %/uA @ 20°C. If the probe slope (1 mil membrane) is greater than 12.6 %/uA or lower than 5.9 %/uA, the membrane/probe may need service. See 12. Troubleshooting, Error Codes, E3 and E4. C: The temperature at the time of calibration in °C. Sensor Press the [SENSOR] soft-key to display the following sensor diagnostics screen: System Parameters Cal. Temperature Probe Slope Probe Offset Probe Current Pressure Offset Sensor = = = = = 22.85 C 6.43 %/uA @20C 0.000 uA 16.91 uA 830 mbar P = 981 Cal.
Reset Ps The [RESET Ps] soft-key (Reset Parameters) is used to reset all instrument parameters to the factory default values. This is useful when the parameters have been changed or corrupted, such as when new software is loaded. Press the [RESET Ps] soft-key to display the following: Reset Press ENTER to RESET Parameters HISTORY SENSOR HELP RESET Ps Press [ENTER] to confirm. All parameters, including calibration values, will be reset.
9. Principles of Operation YSI dissolved oxygen probes use membrane-covered, Clark-type polarographic sensors with built-in thermistors for temperature measurement and compensation. A thin permeable membrane, stretched over the sensor, isolates the sensor elements from the environment, but allows oxygen and certain other gases to enter. When a polarizing voltage is applied across the sensor, oxygen that has passed through the membrane reacts at the cathode, causing a current to flow.
10. Understanding Dissolved Oxygen Error Factors The major sources of error in DO measurement are the accuracy of the instrument components, the accuracy of the probe, and the user’s ability to calibrate the system precisely. Most errors can be reduced substantially by calibrating at DO levels and probe temperatures as close as possible to the expected measurement DO levels and temperatures. In the following, individual sources of error and their ranges are listed.
c. Temperature compensation error: The variation from nominal response to sample temperature is ±0.2% of the DO reading per degree C of the temperature difference between the temperature of the sample and the temperature at which the probe was calibrated. Type 3 Errors: Calibration a. Sample temperature uncertainty error: ±1% of reading This error is approximately zero when using AUTO CAL, when manually calibrating in the % air saturation mode or when calibrating to a Winkler titration sample. b.
Calculation Type Description Calculation Error (mg/L) 1a instrument accuracy (±0.001 x 7.26) + 0.01 ±0.017 2a probe background ±0.01 x (1- (8.24/9.09)) x 7.26 ±0.007 2b probe nonlinearity ±0.003 x 7.26 ±0.021 2c temp. compensation ±(24-20) x 0.002 x 7.26 ±0.058 3a temp. uncertainty ±0.01 x 7.26 ±0.073 3b pressure ±0.012 x 7.26 ±0.087 Probable error can be determined by an r.m.s. calculation: r.m.s. error = [.0172 + .0072 + .0212 + .0582 + .0732 + .0872 ]1/2 = ±0.
11. Maintenance Batteries Low batteries are indicated on the display by “LOW BATT” as shown here. * 8.20 mg /L Main LOW BATT 11:33AM 24.38°C STORE Low battery warning REVIEW SEND CALIBRATE When this happens, replace the batteries with 4 fresh alkaline C-size batteries as soon as possible. • Turn off the instrument. • Turn the instrument over and slide the battery cover to the right (as indicated on the cover) and remove it.
• Remove the old batteries and install fresh alkaline batteries. NOTE: Always observe the correct polarity when installing the batteries. • Replace the battery cover and slide it to the left to lock it into place. NOTE: The instrument parameters are stored in nonvolatile memory and will not be lost during battery replacement.
12. Troubleshooting Error Messages The instrument performs a Power On Self Test each time it is turned on. The following error codes are provided to facilitate troubleshooting.
Code/Problem Error Possible Cause Correction E4: Check Membrane DO High Current Membrane too thin Try another membrane Electrodes need resurfacing Repair or replace Internal leakage in probe or cable connector Repair or replace Short circuit in probe or cable Repair or replace Membrane damaged Replace membrane Probe needs servicing See probe instructions Connector improperly installed Check probe connection Intermittent connection in cable or plug Repair or replace Probe malfunction R
13. Warranty & Repair YSI 5000 Series Dissolved Oxygen Instruments are warranted for two years from date of purchase by the end user against defects in materials and workmanship. YSI dissolved oxygen probes and cables are warranted for one year from date of purchase by the end user against defects in material and workmanship. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determines to be covered by this warranty.
Cleaning Instructions NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic materials must be cleaned and disinfected. Biological contamination is presumed for any instrument, probe, or other device that has been used with body fluids or tissues, or with waste water. Radioactive contamination is presumed for any instrument, probe or other device that has been used near any radioactive source.
Packing Instructions 1. Clean and decontaminate items to insure the safety of the handler. 2. Complete and include the Cleaning Certificate. 3. Place the product in a plastic bag to keep out dirt and packing material. 4. Use a large carton, preferably the original, and surround the product completely with packing material. 5. Insure for the replacement value of the product.
14. Required Notice This equipment generates and uses radio frequency energy and if not installed and used properly, may cause interference to radio and television reception. There is no guarantee that interference will not occur in a particular installation.
15.
16. Appendix A - Oxygen Solubility Table Solubility of Oxygen in mg/l in Water Exposed to Water-Saturated Air at 760 mm Hg Pressure as a function of temperature. Salinity = Measure of quantity of dissolved salts in water. Chlorinity = Measure of chloride content, by mass, of water. S(0/00) = 1.80655 x Chlorinity (0/00) ) Temp o C Chlorinity:0 Salinity:0 5.0 ppt 9.0 ppt 10.0 ppt 18.1 ppt 15.0 ppt 27.1 ppt 20.0 ppt 36.1 ppt 25.0 ppt 45.2 ppt 0.0 14.62 13.73 12.89 12.10 11.36 10.66 1.0 14.
Temp o C Chlorinity:0 Salinity:0 5.0 ppt 9.0 ppt 10.0 ppt 18.1 ppt 15.0 ppt 27.1 ppt 20.0 ppt 36.1 ppt 25.0 ppt 45.2 ppt 21.0 8.92 8.46 8.02 7.61 7.21 6.84 22.0 8.74 8.30 7.87 7.47 7.09 6.72 23.0 8.58 8.14 7.73 7.34 6.96 6.61 24.0 8.42 7.99 7.59 7.21 6.84 6.50 25.0 8.26 7.85 7.46 7.08 6.72 6.39 26.0 8.11 7.71 7.33 6.96 6.62 6.28 27.0 7.97 7.58 7.20 6.85 6.51 6.18 28.0 7.83 7.44 7.08 6.73 6.40 6.09 29.0 7.69 7.32 6.96 6.62 6.30 5.99 30.
17. Appendix B - Pressures and Altitudes Calibration Values for Various Atmospheric Pressures and Altitudes inches Hg 30.23 29.92 29.61 29.33 29.02 28.74 28.43 28.11 27.83 27.52 27.24 26.93 26.61 26.34 26.02 25.75 25.43 25.12 24.84 24.53 24.25 23.94 23.62 23.35 23.03 22.76 22.44 22.13 21.85 21.54 21.26 20.94 20.63 20.35 20.04 19.
18. Appendix C - Conversion Factors TO CONVERT FROM TO EQUATION Feet Meters Multiply by 0.3048 Meters Feet Multiply by 3.2808399 Degrees Celsius Degrees Fahrenheit 9/5×(oC)+32 Degrees Fahrenheit Degrees Celsius 5/9×(oF-32) Milligrams per liter (mg/l) Parts per million (ppm) Multiply by 1 Conversion Factors for Common Units of Pressure kilo Pascals mm Hg millibars inches H20 PSI inches Hg 1 atm 101.325 760.000 1013.25 406.795 14.6960 29.921 1 kiloPascal 1.00000 7.50062 10.
19. Appendix D - Remote Command Language Remote commands and data output are available only when the instrument is in the Remote mode. Command Format: Prefix: The prefix consists of the two character sequence &. In HEX this is 1B 26. Suffix: The suffix consists of the two character sequence . In HEX this is 0D 0A. Command Types and Parameters: Command types are arranged into two groups, immediate and automatic.
The P commands are for requesting system information. Command type parameter description P 0 none Request calibration parameters P 1 none Request solubility table dump P 2 none Request raw A/D P 3 none Request System status The system Setup commands. Command type parameter description S 0 nn.n set salinity to nn.n S 1 hh:mm:ss set time of day, 24 hour clock S 2 mm/dd/yy set calendar S 4 1 emit an audio beep from beeper S 5 0 DO.
1725 Brannum Lane Yellow Springs, Ohio 45387 USA 937 767-7241 • 800 765-4974 • Fax 937 767-9353 Info@ysi.com • www.YSI.