5.6 PERIODIC CALIBRATION
To ensure correct measurement, it is essential that the instrument is regularly calibrated. If used continuously on a
daily basis, a six-month calibration period is recommended, otherwise annual calibration is sufficient. Contact your
distributor for calibration details.
5.7 SERVICE
For other repairs, either under or outside of warranty, please contact your distributor for further information.
5.8 TROUBLESHOOTING
If Esc or Enter button is held pressed while switching on the instrument, the instrument will not start. In that
case, batteries might need to be removed before starting the instrument again. After the batteries have been
reinserted, power on the instrument without holding any keys to start it normally.
If the instrument is switched on on flat batteries and without power supply adapter connected, power supply
circuitry may lock itself. In that case, the instrument will not power up even with external power supply
adapter is connected afterwards. To remedy this situation, batteries must be removed and re-inserted, and
the instrument must be switched on with the power supply adapter. After the successful boot, batteries
should be charged for at least a half of hour to avoid recurrence of this problem on next measurement
session.