Operator`s manual
GAS ANALYZER 361
4. Does not read zero when sampling ambient air 
If the hydrocarbons (HC) reading is above 20 ppm immediately following a zeroing and 
the probe is not measuring vehicle exhaust, this is considered a Hydrocarbon Hangup 
condition. This is caused by HC residues trapped somewhere in the sample system. First 
re-zero the Gas Analyzer. Next, run the analyzer sampling through the sample port with 
the hose off to determine if the residue is in the hose. If the buildup is in the hose, blow the 
hose out with an air source that does not contain oil. Run the analyzer with the hose 
attached for 10 to 30 minutes, keeping the probe free of vehicle exhaust. In addition, check 
the filters for buildup. Replace if dirty. Remember that the lower filter in the Filter Bowl 
gets dirty from the inside out. 
5. NOx readings are erratic 
If the NOx readings bounce around when little or no NOx is present or jump to 1000 or 
more for no reason, most likely the sensor needs to be zeroed and calibrated again. 
6. Leak Test fails 
If the Leak Test fails initially, recheck any filter that may have just been serviced. Next, 
remove the hose, cover the sample port and retest. If it passes, the problem is in the hose. 
Otherwise, separate the filter bowl, plug the inlet, and retest. If it still fails, the problem 
lies inside the Gas Analyzer. 
7. Locks up 
If the Tester locks up, turn off the Tester. Press ON and re-enter the previous test 
environment. If the Gas Analyzer still does not communicate, recycle power to the Gas 
Analyzer. Check for electrical interference sources such as spark plugs and wires, faulty 
regulator, and alternators. 
8. One I/P port works, but the other one does not 
If one I/P port works, but the other one does not work, then the hardware is 
malfunctioning. Both ports are identical and interchangeable. 










