Operator`s manual
GAS ANALYZER 353
Single-Point Calibration 
Press 1 on the <CALIBRATION> menu shown previously. 
Entering new gas values (from Cal Gas bottle) 
The screen at the left is displayed if 1 is selected. Use the u and d keys to move the 
cursor to a new gas value. Use 0 to 9 to enter a new gas value. For example, to enter a 
CO2 value of 11.0%, use the u and d keys to move to CO2 and enter 110 using the 
number keys; 11.0% should be displayed where the cursor is. 
Pressing the r key erases a digit. For example, if you wanted to enter 9.00% for CO, but 
accidentally entered 90.00%, pressing the r would remove the last digit and 9.00% would 
be displayed. 
Pressing the l key changes the value at the cursor location to a 0. 
IMPORTANT: If the calibration gas does 
not contain NOx, set the NOx gas 
concentration to 0 ppm. If the 
concentration is not set to 0 ppm, the unit 
calibrates the NOx sensor incorrectly. As a 
result, the NOx readings will be extremely 
erratic. This can be corrected by 
recalibrating the unit with NOx. 
When you have entered the correct values, 
press E. 
If the Maintenance Required status 
message “Single Pt Cal Warn” appears 
after completion of the calibration, check 
your gas bottle and repeat the calibration 
procedure. If the warning message persists, 
contact Vetronix at (800) 321-4889. Ask 
for Toyota Customer Assistance. 
Calibration Verification 
To verify the correct operation of the Gas Analyzer after a calibration, return to <DATA 
LIST> mode and press A# to shut off the pumps, then feed calibration gas into the 
calibration port of the analyzer. The display should show the gas bottle readings for CO, 
CO2, and NOx. To get the equivalent HC readings as shown on the screen, divide the 
bottle propane concentration by 2 (i.e., 1200 ppm on the bottle will read about 600 ppm on 
the Gas Analyzer).
CO2
CO
Propane
NO
X
12.0%
4.00%
1200 ppm
2000 ppm
FIGURE 15. Entering New Gas Values (from Cal Gas Bottle)










