60061-7 (2021-01) Accula SARS-CoV-2 IFU 9
Insert the tip of the pipette all the way into the sample port of the Test
Cassette until resistance is met.
Squeeze the TOP bulb of the pipette firmly to dispense all of the sample from
pipette stem into the Test Cassette.
NOTE: A small amount of sample may remain in the overflow chamber
(lower bulb). This is normal.
Invert SARS-CoV-2 Buffer vial to mix.
Remove the cap from the eluted patient sample in the SARS-CoV-2 Buffer.
Firmly squeeze the TOP bulb of the pipette.
While continuing to squeeze the top bulb firmly, place the pipette tip well below
the surface of the liquid in the SARS-CoV-2 Buffer vial.
Keep the pipette tip well below the surface of the liquid of the vial containing
the eluted patient sample in SARS-CoV-2 Buffer vial.
Slowly release the top bulb to completely fill the pipette stem with sample.
Some liquid may also be in the overflow reservoir.
NOTE: 
the liquid in the pipette stem will be dispensed.
Completely remove the foil tab covering the sample port on the Test
Cassette. Discard the foil tab.
NOTE: Once the tab is removed from the sample port, sample must be added
immediately (within 5 minutes).
Squeeze
Pipette Stem
Overflow Chamber
DO NOT SQUEEZE