Product Manual
8 60061-7 (2021-01) Accula SARS-CoV-2 IFU
Open the Dock by depressing the black button located on the top left.
Verify the Dock screen displays:
READY INSERT
Do not open the foil pouch until the sample is ready for
testing. Initiate the test within 30 minutes of opening.
Remove a Test Cassette and Transfer Pipette from the foil package (these
items are packaged together).
Write the patient identification (ID) information and testing date on the
Test Cassette label in the area provided.
NOTE: The foil pouch contains a desiccant pack. This can be discarded after Test
Cassette and Transfer Pipette are removed.
Insert the Test Cassette into the Dock, leaving the lid open. Press the Test
Cassette down firmly to seat it in the Dock.
NOTE: Do NOT
remove
the
foil
tab
co
v
e
r
i
ng
the
s
a
m
p
l
e
port
un
t
il
i
mm
e
d
i
a
t
e
l
y
b
e
f
o
r
e
t
e
s
t
i
ng.
Do not close Dock lid until sample has been added to the Test
Cassette.
Verify the Dock screen displays: SARS-COV-2 CASS.
The SAMPLE THEN CLOSE
Once the test cassette is placed into the Dock, you
have 5 minutes to add the sample into the Test
Cassette.