Product Manual
12 60061-7 (2021-01) Accula SARS-CoV-2 IFU
INTERPRETATION OF RESULT
NOTE: LOOK CLOSELY WHEN INTERPRETING RESULTS! T
result that is interpreted as positive for SARS-CoV-2
C = Internal Positive Process Control
T = SARS-CoV-2
NC = Internal Negative Process Control
*If an invalid result is obtained, the sample may be rerun with a fresh Test Cassette only if the eluted sample in Accula buffer has
been stored for less than 2 hours at room temperature. (15°C - 30°C or 59°F -86°F). Alternatively, a new sample should be
collected and run with a new Buffer and Test Cassette.
NOTE: The absence of a Blue Test Line at the position in conjunction with a Blue Test Line at the T position means that the
SARS-CoV-2 target was amplified and detected as a valid result. This can occur due to the overabundance of SARS-CoV-2 target that
competes with the Control target.