Item Brochure

TEST PROCEDURE
STEP 3
Immediately put the swab into the Test Tube.
Vigorously mix the solution by rotating the swab forcefully
against the side of the Test Tube at least ten (10) times.
Best results are obtained when the specimen is vigorously
extracted in the solution.
Let stand for 2 minutes.
STEP 2
Vigorously shake the Extraction Reagent
Bottle 35 times to mix the contents. The
liquid in the Extraction Reagent Bottle
should turn from pink to light yellow.
Add only 6 drops of the Extraction
Reagent to the Test Tube.
STEP 1
Just before testing, squeeze the Extraction Reagent Bottle
to crush the ampule inside. Note: The ampule must be
crushed before proceeding to the next step.
STEP 5
Remove the Test Stick(s) from the container;
re-cap the container immediately.
Place the Absorbent End of the Test Stick into the
extracted sample.
STEP 6
Read results at 5 minutes. Positive results may be read
as soon as the red Control Line appears. Negative results
must be conrmed at 5 minutes.
Results are invalid after the read time.
The use of a timer is recommended.
STEP 4
Express as much liquid as possible from the swab by
squeezing the sides of the tube as the swab is withdrawn.
Discard the swab.
Absorbent End Result Window Handle End
2 min
5 min
10x
6 drops
Note:
Adding more than 6 drops during this step
may provide erroneous results