Product Manual

Open the lid of the Dock.
Remove the Test Cassette and interpret the results according to the
Interpretation of Results section below.
NOTE: Results should be interpreted within 1 hour of test
completion
Dispose of the Test Cassette in the biohazardous waste container.
INTERPRETATION OF RESULTS
Note: The appearance of any shade of Blue Test Line at the FLU A and/or FLU B positions
is a valid result that is interpreted as positive for the influenza A and/or influenza B viral
RNA target. A negative result will only contain a Blue Test Line at the C position.
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
C
FLU A
FLU B
NC
Window
Interpretation
Positive test for
Flu A
Positive test for
Flu B
Positive test for
Flu A and Flu B
Negative test for
Flu A and Flu B
Invalid Results*
The appearance of ANY shade of a Blue
Test Line at the Flu A position indicates a
positive result for the presence of Flu A.
WITH OR WITHOUT the appearance
of a blue process control line at the
C position
AND the absence of a negative process
control line at the NC position
The appearance of ANY shade of a Blue
Test Line at the Flu B position indicates a
positive result for the presence of Flu B.
WITH OR WITHOUT the appearance
of a blue process control line at the
C position
AND the absence of a negative process
control line at the NC position
The appearance of ANY shade of a Blue
Test Line at both the Flu A and Flu B
position indicates a positive result for the
presence of Flu A and Flu B.
WITH OR WITHOUT the appearance
of a blue process control line at the
C position
AND the absence of a negative process
control line at the NC position
The absence of ANY shade of a Blue Test
Line at both the Flu A and Flu B position
indicates a negative result for the presence
of Flu A and Flu B.
AND the appearance of a blue process
line at the C position
AND the absence of a negative process
control line at the NC position
The appearance of ANY shade of a negative
process control line at the NC position,
indicates an invalid test.
The appearance of ALL or NO lines at the C,
FLU A, FLU B, and NC position, indicates an
invalid test.
Window
Window
C
FLU A
FLU B
NC
*If an invalid result is obtained, the sample may be rerun with a fresh Test Cassette only if the prepared sample has been stored
for less than 1 hour at room temperature. (15°C - 30°C or 59°F - 86°F).
NOTE: The absence of a Blue Test Line at the “C” position in conjunction with a Blue Test Line at the “Influenza A,” and or "Influenza B"
position means that the Influenza A and/or Influenza B viral RNA target was amplified and detected as a valid result. This can occur
due to the overabundance of “Influenza A,” and or "Influenza B target that competes with the Control target.
C = Internal Positive Process Control
FLU A = Influenza A
FLU B = Influenza B
NC = Internal Negative Process Control
9 10
TEST COMPLETE
READ RESULTS
P/N: 50033 Rev:0