Product Manual

Remove the patient nasal swab from the Nasal Swab Buffer vial and discard it
into a biohazardous waste container.
Replace the cap on the Nasal Swab Buffer vial.
Write the patient identification (ID) information and testing date onto the Nasal
Swab Buffer vial label in the area provided.
If immediate testing is not possible, recap the Nasal Swab Buffer vial. The
prepared sample may be stored at room temperature (15°C - 30°C, 59°F - 86°F)
for up to 1 hour.
All clinical samples must be at room temperature before beginning the assay.
Check expiration date on each individual Test Cassette foil pouch or outer box before using. Do not use any Test after the
expiration date on the label.
Place Dock on a flat surface.
Connect the AC Adapter to the Power Cord.
Insert round end of the power cord into the Dock. Plug the AC end
of the power cord into an electrical outlet.
Open the Dock by depressing the black button located on the top left.
Verify the Dock screen displays:
“DOCK READY INSERT CASSETTE”.
DOCK READY
INSERT CASSETTE
Do not open the foil pouch until the sample is ready for testing. The Test
must be initiated within 30 minutes of opening the foil package.
Remove a Test Cassette and Transfer Pipette from the foil package (these
items are packaged together).
Write the patient identification (ID) information and testing date on the
Test Cassette label in the area provided.
NOTE: The foil pouch also contains a desiccant pack. This can be discarded
with the foil pouch after Test Cassette and Transfer Pipette are removed.
Insert the Test Cassette into the Dock, leaving the lid open. Press the Test
Cassette down firmly to seat it in the Dock.
NOTE: Do NOT remove the foil tab covering the sample port until
immediately before testing.
Once the test cassette is placed into the Dock, you have 5 minutes
to add the sample into the cassette.
Do not close Dock lid until sample has been added to the Test Cassette.
Verify the Dock screen displays:
“FLU A/FLU B CASS. INSERTED”
The Dock screen will then display:
ADD SAMPLE THEN CLOSE LID”
FLU A/FLU B
CASS. INSERTED
ADD SAMPLE
THEN CLOSE LID
Invert Nasal Swab Buffer vial to mix.
Remove the cap from the prepared patient sample in the Nasal Swab
Buffer and set it aside.
TEST PROCEDURE
DOCK READY
INSERT CASSETTE
DOCK READY
INSERT CASSETE
5 6