Product Manual
8
Refer to the test specific Instructions For Use for sample collection and preparation.
The Silaris Test Cassettes are for use with the Silaris Dock Only
Insert the Test Cassette into the slot with the cassette label facing toward
the front of the Dock.
NOTE: The sample port (covered with a foil tab) on the top of the cassette
should be accessible to the user.
Fully depress the Test Cassette into the Dock until it ‘clicks’ into position.
The Silaris Dock will measure the temperature of the cassette and display it.
NOTE: Once cassette is seated into the Dock, the user will have only
5 minutes to start the test before the Dock times out.
If the cassette has been properly inserted, the display will identify the
type of test that is supported by the cassette.
If the display shows an error code, refer to
Section 12 -Troubleshooting for additional information.
Remove the Test Cassette from the foil pouch by tearing off the top of the
pouch starting at the notches in the pouch.
T
E
S
T
C
O
M
P
L
E
T
E
R
E
A
D
R
E
S
U
L
T
S
9. DOCK OPERATION
This section provides detailed instruction pertaining to the operation of the Silaris Dock only. It is not intended to be a comprehensive
i
nstruction for performing diagnostic tests. Such instructions are found packaged with the specific Silaris Test Kits.
The Silaris Dock is capable of running a number of Silaris Tests. For the purpose of illustration, this section includes Dock messages
specific to the Silaris Influenza A&B Test.
9.1 INSERTING THE TEST CASSETTE
Each Test Cassette may only be used once. Sensors in the Dock will not allow the assay to run if the cassette has
already been inserted.
CASSETTE
TEMPERATURE: 24C
After a few seconds the display will prompt the user to add the sample
to the cassette sample port.
The Test cassette is now ready to for the addition of the sample.
To avoid contamination, do not remove the foil tab covering the
cassette sample port until you are ready to add the sample.
Open the lid of the Dock by depressing the black lid latch button and
pulling the lid to the open position.
FLU A/FLU B
CASS. READY
ADD SAMPLE
THEN CLOSE LID