Product Manual

5
7.1 EXTERIOR CLEANING PROCEDURE
Cleaning of the exterior surfaces of the Silaris Dock and the immediate work area should be performed as necessary, no less than daily
when in use.
1
. Unplug the power cord from the wall outlet and Dock.
2. Close the Dock lid.
3. Using a cleaning solution (10% bleach or 70% ethanol), or a germicidal disposable wipe (such as CaviWipes™), gently wipe the
outer surfaces of the Dock, removing any debris.
NOTE: Do not press the wipe against the open vents of the back or sides of the Dock.
4. Using a new cloth, wipe the front of the Dock twice top to bottom, then twice left to right. Follow this procedure for the back, top
and bottom of the Dock.
5. Do not let liquid accumulate near any opening. Make sure no liquid enters the dock.
6. Leave the Dock “wet” for at least 3 minutes.
7. With a fresh dry cloth (such as a Kimwipe™) remove any residual moisture
8. Allow the unit to dry for at least 10 minutes prior to re-connecting the power cord for the AC Adapter.
7.2 DISPOSAL PROCEDURE
All electronic products must be disposed of separately from the municipal waste system.
A used Dock may be biologically contaminated. Prior to disposal, or shipping for services, the Dock must be
thoroughly cleaned.
The Silaris Dock should be disposed in appropriate electronic recycling waste. For more information regarding
disposal of the Dock, please contact Sekisui Diagnostics Technical Support at 800-332-1042 (U.S Only)
or 781-652-7800 (Outside U.S).
To decontaminate the Dock prior to disposal:
1. Unplug the DC power connector from the back of the Silaris Dock.
2. Open the lid of the Dock by depressing the lid latch button and pulling the lid to the open position.
3. If there is a Cassette inserted into the Dock, remove it by firmly grasping and pulling it upward.
4. Soak the entire Dock in a 10% bleach solution for at least 10 minutes. Ensure that the Dock is fully submerged in the solution,
and that the entire interior of the Dock is flooded. It is recommended to maintain an upright orientation of the Dock in order to
allow entrapped air to escape through the cassette slot.
5. Remove the Dock from the bleach solution. Turn the Dock upside down to drain the solution from the interior. Allow the Dock to dry.
6. Dispose of the Dock as electronic waste in accordance to local regulations.
NOTE: If for any reason, it becomes necessary to return a Dock to Sekisui Diagnostics, prior to packaging please carefully
follow the exterior cleaning procedure. Do not follow this disposal Decontamination Procedure.