Product Manual
6. SPECIFICATIONS
6.1. DOCK
4
P
OWER INPUT 12VDC ± 1VDC, 1A maximum
OPERATING ENVIRONMENT Ambient Temperature: 15°C to 30°C (59 to 86°F)
Relative Humidity: 20%-80%, non-condensing
Altitude: Sea level to 8,000 feet
DIMENSIONS 5.7 in (14.5 cm) H x 3.9 in (9.9 cm) W x 3.8 in (9.7 cm) D
TOTAL WEIGHT
(Dock, Power Supply, and Power Adaptor) < 1.4 lbs (619 g)
ACCEPTIBLE TILT The Dock must be installed on a level surface, with an incline not to exceed
10 degrees.
The Silaris Dock is not to be used in direct sunlight.
OPERATIONAL LIFE 1,500 tests
6.2. AC ADAPTER
POWER INPUT 100-240VAC, 50/60Hz
POWER OUTPUT 12VDC, 2.08A, 25W maximum
NOTE: Use only the supplied power adapter with the Silaris Dock. This component has been tested and certified for use.
If a replacement power supply is needed, please contact Sekisui Diagnostics Technical Support for assistance at
800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S).
7. DOCK MAINTENANCE
To reduce the risk of damage to the Dock, avoid use of sprays in the vicinity of the Dock.
Prior to cleaning the Silaris Dock, disconnect the power supply. This will help to prevent damage to the unit.
Ensure that the Dock lid is closed prior to cleaning.
Do not open the Dock lid during the cleaning procedure.
Do not insert the cleaning wipe into the cassette slot. This may cause damage to the unit.
Do not use solvents other than those recommended above.
Do not let liquid accumulate in or around Dock vents
Regular cleaning of the Silaris Dock is recommended. The Dock should be cleaned every day that it is used, and prior to disposal or
shipment or servicing.