User manual
Maintenance
TPG User Manual 71
3. Click the icon .
The current parameter values are displayed.
4. Save the 'parameters' file on a local system with the help of
your browser.
ª The parameter file is copied and secured.
Loading the parameters file to a TPG
Proceed as follows:
1. Start the TPG Control Center.
2. Select MAINTENANCE - Parameter backup.
3. Click Browse.
4. Specify the 'parameters' file.
5. Click Import.
ª The parameter values in the file are applied to the TPG.
You can also automatically load a parameters file from a USB flash
drive to a TPG; see: Ö71.
6.2 How to Use a Connected USB Device
You can connect a USB flash drive to the USB port of the TPG to
make use of additional features of the TPG.
Print job buffering
If a printer is not available, the print jobs that are sent to that
printer can be automatically buffered in the USB flash drive. As soon
as the printer becomes available, the print jobs will be transferred to
the printer and printed. If this option has been disabled and the
printer is not available, the print jobs will be discarded after a
specified timeout (’printer connection timeout’ Ö44).
Parameter Backup
During the 'parameter backup', the 'parameters' file will be saved
automatically on the USB flash drive and updated after a parameter
change. The file contains all parameter values of the TPG (exception: