Specifications
Operating Manual SEFRAM 4451
M4451 A00
69
input with a 50Ω termination. Select leading and
trailing times.
CHECK - The Rise and Fall times measured between 10% and
90% points are within specification.
Amplitude Accuracy Test
SPECIFICATION - ± (1% of setting +10mV)
into 50Ω.
PROCEDURE - Connect the instrument OUTPUT to the oscilloscope
input with a 50Ω termination. Change the high and low
levels to obtain different signal amplitudes.
Select a period of 100us.
CHECK - The signal amplitude measured is within specification.
Internal Trigger Accuracy Test
SPECIFICATION - ± 0.01% from 100ns to 100s.
PROCEDURE - Connect the instrument SYNC OUT to the counter
input A, set to measure period in Auto Trig mode.
Select INT TRIGGER mode on SEFRAM 4451 and change
the rate to different values.
CHECK - The trigger rate measured by the counter is within
specification.
GPIB Capability
Connect the SEFRAM 4451 generator to a controller. Each side of this connection must be fitted with a
proper IEEE-488 interface. Operate the instrument via the GPIB bus and change setting as desired. This
procedure assumes that the operator has some experience in programming the controller and is familiar with
the GPIB standard.