Specifications

Operating Manual SEFRAM 4451
M4451 A00
46
:DOUBle
[:STATe] <Boolean>
:DELay <NRf>
:TRANsition
[:LEADing] <NRf>
:TRAiling <NRf>
:AUTO <Boolean>|ONCE
:POLarity NORMal|COMPlement|INVerted
4.12.1.1 Frequency
The frequency command controls the frequency of the pulse in the continuous trigger mode. It
is the inverse of the period.
COMMAND TYPE: Setting or Query
SETTING
Syntax: [:SOURce]:FREQuency[:CW|FIXed]<ws><frequency>[units]
Arguments:
Type: NRf
Units: MHz, kHz, Hz (default)
Range: 0.1Hz to 50MHz
Rounding: To the resolution of the range.
Examples: :FREQ 5KHZ
:FREQ 5E3
QUERY
Syntax: [:SOURce]:FREQuency[:CW|:FIXed]?
Examples: :FREQ?
Response: NR3
CONSIDERATIONS: FIXed is an alias for CW.
4.12.1.2 High Voltage Level
This command is used to set the high level of the pulse.
COMMAND TYPE: Setting or Query
SETTING
Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:HIGH<ws><high level>[units]
Arguments:
Type: NRf
Units: MV, V (default)
Range: -9.5V to +10V
Rounding: To 10mV
Examples: VOLT:HIGH 4V
QUERY
Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:HIGH?
Examples: :VOLT:HIGH?
Response: NRf
CONSIDERATIONS:
1) The high level must be greater than the low level.
2) The difference between the levels must conform to 0.5V difference 10V
3) The high level may not exceed the high limit.
4.12.1.3 Low Voltage Level