Specifications

Operating Manual SEFRAM 4451
M4451 A00
28
Trigger Menu
F1: MAN - Selects manual as the trigger source. Pressing the MAN TRIG key generates
the
trigger. In the Gate trigger mode, the pulse is generated as long as the key is being pressed.
F2: INT - Selects the internal trigger generator as the trigger source. Change the internal
trigger rate displayed with the rotary input knob or numerical keys. The rate has a range of
100ns to 99.99s, although the minimum value is limited by the value of the period in that
the rate cannot be less than the period.
Internal Trigger
F3: EXT - Selects the external trigger signal as the trigger source. The trigger source is
supplied through the TRIG IN connector.
F4: LEVEL/SLOPE - Two parameters are related to external trigger source operation.
These are LEVEL and SLOPE. The Level determines at what voltage level the external
signal will be recognized as a trigger. At level less that this, no pulse will be generated. The
Slope determines whether the positive or negative edge of the trigger signal will trigger the
pulse
3.6.5 SETUPS Menu
The pulse generator can store the current front-panel settings, called a setup, into one of 99 storage buffers.
When you recall a setup, the pulse generator restores the front-panel settings to those that you stored in the
selected buffer. Because it is impossible to 100% guarantee against loss of stored data, you should maintain
a record of the data stored in memory so that you can manually restore such data, if necessary.